Electrolab RediLevel 2100 DLS Скачать руководство пользователя страница 10

RediLevel 2100 DLS  

 

 

 

 

 

 

 

 

                          page 6 

EL 29306    

 

 

 

 

 

 

 

© 2022  Electrolab, Inc. All Rights Reserved.

 

Assign Unit Number to Sensor With the Corresponding Serial Number (version 

3.15 and higher) 

UsssssssNnn 

sssssss = seven-digit serial number 

nn = two-digit unit number 

 

Response: 

UuuNOK

    uu = new unit number 

UuuEEerr  

 Write to EEPROM failed

 

Set Baud Rate 

UuuBbbbb[b][pds] 

Where 

bbbb[b] = 1200, 9600, 14400, 19200, 38400, or 57600 (9600 is default) (v3.15 and higher) 

pds = parity, data length, stop bit (pds options) 

N81 (default) 

E71  

O71 

For example

to program Unit 00 to 9600 E 7 1 would be U00B9600E71 

 

Note

: It is not necessary to power down the Model 2100 before this command takes effect. The Model 

1000 must have power cycled for this command to take effect. 

 

Response: 

UnnBOK 

Set Number of Floats 

UuuFn 

Where 

uu = unit number 

n = float number (1 or 2 – Standard; 11 or 12 – 1/8” Resolution; 11 – Single Float; 12 – Dual Float) 

 

Response: 

UuuFOK 

Set Level Offset 

UuuLOslll 

Sets the offset for the level sensor 

Where 

uu = unit number 

sIIlI = sign and offset with two implied decimal places 

Note:

 If two floats, assigns the same offset to both 

 

Response: 

UuuOLOK 

Set Level Offsets for Individual Floats 

UuuL[1|T]Oslll.ll

  Sets the top float offset. 

Example of setting top level offset for Unit 00 to 2.25 inches - U00L1O2.25 

 

UuuL[2|B]Oslll.ll

 Sets the bottom float offset. (

Dual Float Sensor only) 

Example of setting bottom level offset for water interface to 1.75 inches - U00L2O1.75 

Where 

uu = unit number 

Содержание RediLevel 2100 DLS

Страница 1: ...EL 29306 2022 Electrolab Inc All Rights Reserved USER MANUAL MODEL 2100 DLS PRODUCT LINE REDILEVEL 2100 DLS ...

Страница 2: ...ce with the terms andconditions of the License Agreement between the user and Electrolab Inc Electrolab Inc makes no representations of warranties with respect to the contents hereof and specificallydisclaims any implied warranties of merchantability of fitness for any particular purpose Electrolab Inc reserves the right to revise this publication and to make changes from time to time in its conte...

Страница 3: ...n Unit Number to Sensor With the Corresponding Serial Number version 3 15 and higher 6 Set Baud Rate 6 Set Number of Floats 6 Set Level Offset 6 Set Level Offsets for Individual Floats 6 Set Temperature Offset 7 Set Multiple Temperature Sensor Offsets 7 Set Receive to Transmit Delay 7 Set 4 20mA Poll Period 7 Set 4 20mA Minimum 4mA Range 7 Set 4 20mA Maximum 20mA Range 8 Set the Level Error settin...

Страница 4: ...ponding to Serial Number version 3 15 and higher 10 Report sensor Health Status version 3 15 and higher 10 Report the Level Error Setting version 3 09 and higher 11 Report the Modbus 16 bit Unsigned Integer 32 bit or 2 x 16 bit floating point mode 11 Report Battery Voltage 11 Modbus Registry Map 12 Table 1 Read Write Registers 12 Table 2 Holding Registers 15 Table 3 Holding Registers 2 x 16 bit 16...

Страница 5: ...mmunications require a 2 pair cable Wiring Connections The sensor requires a power supply of 5 6 to 13 volts DC To connect the sensor communication lines to the telemetry equipment connect the sensor data receive to the RTU data transmit and the sensor data transmit to the RTU data receive The voltage supply can be connected to a switched output so power is applied only during sensor polling Insta...

Страница 6: ...vel offset is always added to the raw value If a mistake is made when entering the level offset reset the offset value to zero before proceeding to avoid large swings in readings If there is an offset programmed in the sensor and the actual level is not correctly displayed simply changing the offset value will not include the previous offset value Example If a 1 50 offset is entered and the sensor...

Страница 7: ...There is insufficient voltage The sensor needs at least 5 6 VDC Sensor has been in service for some time but is not working 1 There are corroded connections or damaged cables 2 Blown fuse or shorted suppressor on the barrier board if equipped Check the fuses with a continuity tester 3 There is possible damage to sensor electronics Sensor sends inaccurate level or temperature 1 An incorrect level o...

Страница 8: ...card for either digit Note Do Not use the wildcard if connected to more than one level sensor as all sensors will respond simultaneously Response UuuDlll llFtttEeeeeWwwww where uu unit number lll ll level in inches repeated for sensors with 2 floats ttt temperature in degrees F eeee error number 0 No errors 1 No float detected 2 One float is out of range on a two float sensor 3 Too many groups 4 r...

Страница 9: ... out of range on a two float sensor 3 Too many groups 4 reserved 5 Transmit to slave processor for level failed 6 Transmit to slave processor for temperature failed 7 Receive from slave processor of level failed 8 Receive from slave processor of temperature failed 9 No slave processors responding Note If errors 5 through 9 occur and persist after power cycling the sensor should be returned for rep...

Страница 10: ...re this command takes effect The Model 1000 must have power cycled for this command to take effect Response UnnBOK Set Number of Floats UuuFn Where uu unit number n float number 1 or 2 Standard 11 or 12 1 8 Resolution 11 Single Float 12 Dual Float Response UuuFOK Set Level Offset UuuLOslll Sets the offset for the level sensor Where uu unit number sIIlI sign and offset with two implied decimal plac...

Страница 11: ...ets UuuTnOso o Where uu unit number n temperature sensor identifier 1 8 1 is at top so o 9 9 to 9 9 degrees optional sign Response UuuTnOOK Set Receive to Transmit Delay UuuRmmm Where uu unit number mmm milliseconds 50 to 250 The default is 127ms Response UuuROK Set 4 20mA Poll Period UuuMPpppp Where uu unit number pppp seconds Default is 30 seconds Response UuuMPOK Note Poll Periods less than 20 ...

Страница 12: ...ault setting xxxx 1008 will set the 32 bit floating point mode xxxx 1009 will set the 2x16 bit floating point mode v3 14 and higher Response UuuIFOK Force sensor to enter Boot Load mode version 3 15 and higher future use UUUFB uu two digit unit number Response none Enter High Level Electronic Shut Down ESD mode version 3 15 and higher UuuESDONn uu unit number n one digit number from 1 to 3 which r...

Страница 13: ... Offset UuuOF Response UuuOFsff Where uu unit number s sign ff temperature offset degrees F Report Multiple Temperature Offsets UuuTO Response UuuTnOso o TnOsn n repeated for additional temperature sensors Where uu unit number n temperature sensor 1 8 1 is top sensor s sign o o temperature offset Report Switch Distance UuuD Response UuuDd Where uu unit number d distance between switches as integra...

Страница 14: ... digit serial number Response UsssssssNuu uu unit number Report sensor Health Status version 3 15 and higher UuuH uu two digit unit number Response grp0 ttt bbb grp1 ttt bbb grp3 ttt bbb grp4 ttt bbb UuuDlll ll Dlll ll Fttt Fttt EeeeeWwwww BATTERY VOLTAGE vv vV If sensor works in normal operation parameters the message will be UuuSENSOR OK Error one or more of the following Error messages NO SWITC...

Страница 15: ...vel error report to be 999 99 default x 1 is set for level error report to be 000 00 Report the Modbus 16 bit Unsigned Integer 32 bit or 2 x 16 bit floating point mode UuuIF Response UuuIF x Where uu unit number x 0 is set for 16 bit Unsigned integer mode This is the default setting x 1 is set for 32 bit floating point mode x 2 is set for 2x16 bit floating point mode v3 14 and higher Report Batter...

Страница 16: ...floating point 2 40108 107 1 Factory setting 16 bit unassigned integer see Note for advanced users 16 bit Unsigned Set baud rate 1200 9600 14400 19200 38400 57600 V3 15 and higher 40109 108 1 Factory setting 9600 16 bit Unsigned Set parity 78 N No parity 79 O Odd parity 69 E Even parity 40110 109 1 Factory setting 78 N 16 bit Unsigned Set data bit 8 40111 110 1 Factory setting 8 16 bit Unsigned Se...

Страница 17: ...1 Factory setting 0 16 bit Signed Sensor Description Registers Serial number high 40127 126 1 Read only 16 bit Unsigned Serial number medium high 40128 127 1 Read only 16 bit Unsigned Serial number medium low 40129 128 1 Read only 16 bit Unsigned Serial number low 40130 129 1 Read only 16 bit Unsigned Version number 40131 130 1 Read only 16 bit Unsigned Number of modules 1 to 8 40132 131 1 Read on...

Страница 18: ... Group 1 bottom 40140 139 1 Read only 16 bit Unsigned v3 15 and higher Group 2 top 40141 140 1 Read only 16 bit Unsigned v3 15 and higher Group 2 bottom 40142 141 1 Read only 16 bit Unsigned v3 15 and higher Group 3 top 40143 142 1 Read only 16 bit Unsigned v3 15 and higher Group 3 bottom 40144 143 1 Read only 16 bit Unsigned v3 15 and higher ...

Страница 19: ...g point Water Volume bottom level x K_factor 43996 3995 1 Read Only Water Volume 16 bit Unsigned integer or 32 bit floating point Temperature1 43997 3996 1 Read Only Temperature1 16 bit Unsigned integer or 32 bit floating point Temperature2 43998 3997 1 Read Only Temperature2 16 bit Unsigned integer or 32 bit floating point Temperature3 43999 3998 1 Read Only Temperature3 16 bit Unsigned integer o...

Страница 20: ...top to bottom 45005 5004 2 Read Only Oil Level in Tank top bottom Floating point upper two bytes v3 14 and higher 45006 5005 Read Only Oil Level in Tank top bottom Floating point lower two bytes v3 14 and higher Total Volume top level x K factor 45007 5006 2 Read Only Total Volume top level x K factor Floating point upper two bytes v3 14 and higher 45008 5007 Read Only Total Volume top level x K f...

Страница 21: ...ad Only Temperature6 v3 14 and higher Floating point upper two bytes 45024 5023 Read Only Temperature6 v3 14 and higher Floating point lower two bytes Temperature7 45025 5024 2 Read Only Temperature7 v3 14 and higher Floating point upper two bytes 45026 5025 Read Only Temperature7 v3 14 and higher Floating point lower two bytes Temperature8 45027 5026 2 Read Only Temperature8 v3 14 and higher Floa...

Страница 22: ...e from slave processor for temperature failed Bit 8 256 No slave processor responding Notes for advanced users Writing 1007 at register address 107 will set register 107 to 0 This will set the device registers that hold top float bottom float and temperature to 16 bit Unsigned integer Writing 1008 at register address 107 will set register 107 to 1 and will set the device hold registers 43991 3990 ...

Страница 23: ...ature sensor standard up to 8 temperature sensors available with desired spacing optional 1 5 C accuracy Power Requirements 5 6 VDC to 13 VDC Power Consumption 15mA nominal 20mA maximum Pressure 40 psi standard 250 psi optional Communication RS485 Two or four wire communications Baud rate and parity programmable up to 57600 baud v 3 15 and higher 4 20mA signal available when connected to digital t...

Страница 24: ...wire communication two twisted pair 16 18 AWG recommended Four wire communication three twisted pair 16 18 AWG recommended Classification Class I Div 1 Group D Hazardous Locations when connected to an approved intrinsically safe barrier board Certification ANSI UL 913 CAN CSA C22 2 No 157 ...

Страница 25: ...RediLevel 2100 DLS page 21 EL 29306 2022 Electrolab Inc All Rights Reserved Figure 1 RediLevel 2100 DLS ...

Страница 26: ...RediLevel 2100 DLS page 22 EL 29306 2022 Electrolab Inc All Rights Reserved Figure 2 DLS2100 Assembly Diagram ...

Страница 27: ...RediLevel 2100 DLS page 23 EL 29306 2022 Electrolab Inc All Rights Reserved Figure 3 DLS2100 Connection Diagram ...

Страница 28: ...is 1 longer than the tank height to accommodate the extra height the dome adds 2 Baud Rate and Parity desired Default is 9600 N 8 1 If other baud rate and parity are required please specify 3 Unit Numbers Required If ordering more than one sensor for a location then you may have the level sensors pre addressed with the required unit numbers prior to shipment for the tanks on that location Example ...

Страница 29: ... All Rights Reserved Contact Information For further information or for assistance please contact Electrolab Inc 159 Enterprise Parkway Boerne Texas 78006 Phone 210 824 5364 TF 888 301 2400 Email InsideSales electrolabcontrols com www electrolabcontrols com ...

Отзывы: