RediLevel 2100 DLS
page 6
EL 29306
© 2022 Electrolab, Inc. All Rights Reserved.
Assign Unit Number to Sensor With the Corresponding Serial Number (version
3.15 and higher)
UsssssssNnn
sssssss = seven-digit serial number
nn = two-digit unit number
Response:
UuuNOK
uu = new unit number
UuuEEerr
Write to EEPROM failed
Set Baud Rate
UuuBbbbb[b][pds]
Where
bbbb[b] = 1200, 9600, 14400, 19200, 38400, or 57600 (9600 is default) (v3.15 and higher)
pds = parity, data length, stop bit (pds options)
N81 (default)
E71
O71
For example
,
to program Unit 00 to 9600 E 7 1 would be U00B9600E71
Note
: It is not necessary to power down the Model 2100 before this command takes effect. The Model
1000 must have power cycled for this command to take effect.
Response:
UnnBOK
Set Number of Floats
UuuFn
Where
uu = unit number
n = float number (1 or 2 – Standard; 11 or 12 – 1/8” Resolution; 11 – Single Float; 12 – Dual Float)
Response:
UuuFOK
Set Level Offset
UuuLOslll
Sets the offset for the level sensor
Where
uu = unit number
sIIlI = sign and offset with two implied decimal places
Note:
If two floats, assigns the same offset to both
Response:
UuuOLOK
Set Level Offsets for Individual Floats
UuuL[1|T]Oslll.ll
Sets the top float offset.
Example of setting top level offset for Unit 00 to 2.25 inches - U00L1O2.25
UuuL[2|B]Oslll.ll
Sets the bottom float offset. (
Dual Float Sensor only)
Example of setting bottom level offset for water interface to 1.75 inches - U00L2O1.75
Where
uu = unit number