28
en
17.2 Transducer
The transducer will wear with repeated use. Transducer life depends on the number of
measurements taken and the manner in which readings are taken. To extend transducer life,
always set the transducer down so that it is perpendicular to the panel surface. Dragging the
transducer along the surface will reduce the life of the transducer. Replacement transducers are
available from your local Elcometer NDT supplier or directly from Elcometer NDT.
18 TECHNICAL SPECIFICATION
Measurement Rate
Manual
4 readings per second
Scan mode
32 readings per second
Measuring Range
a
- Sound Velocity
a.
Measuring Range depends on material, surface conditions and the transducer selected.
1250 m/s to 10000 m/s (0.0492 in/µs
to 0.3930
in/µs)
Measuring Range -
Thickness
Pulse-Echo
P-E
0.63 mm to 254 mm (0.025” to 9.999”)
Echo-Echo
E-E
1.27 mm to 102 mm (0.05” to 4.0”)
Range will vary +/- depending on the coating.
Measurement Resolution
Sound
Velocity
1 m/s (0.0001 in/µs)
Thickness
0.01 mm (0.001")
Velocity Calibration Range
1250 m/s to 10000 m/s (0.0492 in/µs
to 0.3937
in/µs)
Weight (including batteries)
380 g (13.5 oz)
Dimensions (W x H x D)
63.5 mm x 165 mm x 31.5 mm (2.5” x 6.5” x
1.24”)
Gauge Operating Temperature
-10°C to 60°C (14°F to 140°F)
Case
Aluminium case with gasket sealed end caps and
waterproof membrane keypad
PC Connection
RS232 serial port. Windows PC interface
software
Display
Backlit VGA greyscale display (240 x 160 pixels).
Viewable area 62 mm x 45.7 mm (2.4” x 1.8”). EL
backlit (on/off/auto).
Power Source
Three 1.5 V AA alkaline or rechargeable cells.
Typically operates for 200 hours on alkaline and
120 hours on rechargeable cells (charger not
included.)
Note: Alkaline batteries must be disposed of
carefully to avoid environmental contamination.
Please consult your local environmental authority
for information on disposal in your region. Do not
dispose of any batteries in fire.