9 TECHNICAL SPECIFICATION
Chemical
Measuring range:
0.1 µg.cm
-2
to 20 µg.cm
-2
soluble salt (as
sodium chloride) when using recommended
High Purity Sample Papers and water with a
conductivity of <0.25 microsiemens
Resolution:
0.1 µg.cm
-2
(with sodium chloride)
Measuring accuracy:
±10% of reading
Temperature correction:
Calibrated at 25°C
Below 25°C: -1.7% /°C
Above 25°C: +1.7% /°C
Electrical
Electrodes:
Copper
Display:
20 digit LCD
Sampling timer:
2 minutes
Battery:
9 V Alkaline PP3 (6LR61)
Timeout:
20 seconds
Working limits
Temperature range:
5°C to 40°C
Maximum relative humidity:
80%
Physical
Dimensions:
200 mm x 190 mm x 60 mm
Weight:
1.5 kg
Case:
Damage resistant polyurethane
Sealing:
Splash proof
Sampling capability
Substrate:
Metallic or non-metallic
Area required:
110 mm diameter
Position:
Any horizontal, vertical, curved or inclined
surface where sample paper can be placed