64
en
DISP
(Display)
VIEW
RF, +RECT, -RECT or RECTFW.
DELAY
Adjust where the left side of the display window starts
according to thickness, in inches or millimetres.
RANGE
Set the overall depth of the viewable measurement area.
It functions a lot like a zoom on a camera.
UNITS
Toggle between Metric (millimetres) or English (inches)
units.
BRIGHTNESS
Adjusts the brightness of the display.
Scale is 0-20 where 20 is the brightest.
SCREEN
UPDATE
Toggles the display refresh raate (60Hz or 120Hz).
The 60Hz option conserves battery life.
COLORS
Choose the display colour scheme (from a choice of 12).
DIM
Conserve battery life by dimming the display after idle for
a specific amount of time: OFF, 30, 60, 90, 120 seconds.
Pressing any key will restore the screen brightness.
RECT WAVE
Toggles between an outlined or filled view option when
the display setting is in RECT (rectified) wave mode.
DETECT MARK
Choose the format of the vertical line on the RF and
RECT measurement screens which shows the detection
point:
Line, Box, Dots, None.
TUNE
FILTERS
The gauge is a wide band flaw detector with a frequency
range of 1.8 MHz to 19 MHz (-3dB). The gauge also has
selectable narrow band filters at 2 MHz, 5 MHz and 10
MHz.
GAIN
Increases or decreases the overall amplitude of the
signal.
Gain is similar to the volume control on a stereo receiver.
DETECT MODE
Sets the detection mode; PEAK, FLANK or Z-CROSS
(zero crossing)
Menu (Flaw Det.)
Function