background image

ELCART

ELCART DISTRIBUTION SPA  via Michelangelo Buonarroti, 46 - 20093 Cologno Monzese (Milano) ITALY

Tel. +39 02.25117310 Fax +39 02.25117610 sito internet:  www.elcart.com    e-mail: [email protected]

Manuale di istruzioni/Scheda tecnica

La divulgazione dei dati contenuti in questa scheda è da ritenersi un servizio puramente informativo e non costituisce alcun vincolo da parte della Elcart in merito a prestazioni ed utilizzo del prodotto.

The divulgation of data contained on this technical sheet are exclusively for informational reasons and establish no link on behalf of Elcart regard to thr performances and the usa of the product.

La divulgacion de los datos contenidos en esta ficha son un servicio unicamente informativo y no constituyen ningun vinculo de parte de Elcart respecto a las prestaciones y uso del producto.

ART. 09/07800-00

PAGINA 3 DI 4

1

7

6

4

3

2

5

1.  LCD Display: 3 ½ digits, character 15mm high

2.  Light Button Switch> Press this button to swtch on back light if the dark  

  circumstance light makes the reading difficulty when measuring, the   

  light will be automatically turned off in 5 seconds. Press again to switch  

  it on again. If the battery is in weak power, the light will be dimmed.

3.  Rotary Switch: use this switch to select functions and ranges 

4.  VΩmA Input Jack

5.  COM Input Jack

6.  Ω10A Input Jack

7.  Data-hold Switch (HOLD)

3. Safety Information

3-1 The meters are designed according to IEC-61010 concerning electronic 

 

measuring instruments with an over-voltage category (CAT III) and   

 

pollution 2.

3-2 Follow all safety and operating instructions to ensure that the meter is  

 

used safely and is kept in good operating condition.

3-3 Safety symbols:

 

Important safety information, refer to the operating manual.

   Dangerous voltage may be presence.

   Double insulation (protection Class II)

4. Special Cautions for Operation

4-1 The meters can be safe only according to standard procedures when   

 

used in conjunctions with the supplied test leads. To replace damaged  

 

test leads with only the same model or same electric specifications.

4-2 To avoid risk of electric shock, do not use the meters before the cover   

 

is in place.

4-3 The range switch should be right position for the testing.

4-4 To avoid electric shock and damaging the instruments, the input  

 

 

signals are forbidden to exceed the specified limits.

4-5 When measuring TV set or switched power, attention should be paid    

 

to the possible pulses that may bring destruction to the circuit.

4-6 Range switch position is forbidden to be changed at random during    

 measurement.

4-7 Take caution against shock in the course of measuring voltage higher   

 

than DC 60V & AC 30V.

4-8 Protection fuse should be replaced only with same type and same  

 

 specification.

5. GENERAL SPECIFICATIONS

5-1 Max Voltage between input terminal and Earth Ground: CAT III 600V 

5-2 Over-range Indication: display “1” for the significant digit.

5-3 Automatic display of negative polarity “-” .

5-4 Low Battery Indication: ‘      ’ displayed

5-5 Max LCD display: 1999 (3 ½ digits)

5-6 Fuse protection:  

F-250mA/600V (Ø5x20mm)

5-7 Power Supply:  

9V battery, 6F22 or NEDA 1604

5-8 Operating temperature:  

0°C to 40°C (relative humidity <85%)

1. Overview 

The multimeter is characterized at slim size, portable, stable performance 

and anti-dropping capacity. Using 3 ½ digits LCD monitor with character 

15mm high, they offer clear readings.  With overall circuitry design centering 

on large-scale IC A/D converters in conjunction and over-load protection 

circuit, the meters give excellent performance and exquisite making as a 

handy utility instrument.

The  meters  can  be  used  to  measure  AC  &  DC  voltage,  DC  current, 

resistance, positive diode voltage fall and continuity.

2. Panel Layout 

DIGITAL MULTIMETER ART. 09/07800-00 NI2100

OPERATION MANUAL

5-9   Storage temperature: 

-10°C to 50°C ((relative humidity <85%)

5-10  Guaranteed precision temperature:  23±5°C (relative humidity <85%)

5-11  Dimensions:  

69x138x31mm (with holster)

5-12  Weight:  

approx 170gr. (including battery)

6. Testing Specifications 

Accuracy is specified for a period of year after calibration and at 18°C to 

28°C (64°F to 82°F) with relative humidity to 75%.

6-1  DC Voltage

Range

Resolution

Accuracy

200mV

0,1mV

±(0.5% of rdg ± 3 digits )

2V

1mV

±(0,5% of rdg ± 3 digits)

20V

10mV

±(0,5% of rdg ± 3 digits)

200V

100mV

±(0,5% of rdg ± 3 digits)

600V

1V

±(0,8% of rdg ± 3 digits)

Range

Resolution

Accuracy

20µA

0,01µA

±(1,0% of rdg ± 3 digits)

200µA

0,1µA

±(1,0% of rdg ± 3 digits)

2mA

1µA

±(1,0% of rdg ± 3 digits)

20mA

10µA

±(1,0% of rdg ± 5 digits)

200mA

100µA

±(1,5% of rdg ± 5 digits)

10A

10mA

±(3,0% of rdg ± 10 digits)

Range

Resolution

Accuracy

200Ω

0,1Ω

±(0,8% of rdg ± 5 digits)

2KΩ

±(0,8% of rdg ± 2 digits)

20KΩ

10Ω

±(0,8% of rdg ± 2 digits)

200KΩ

100Ω

±(0,8% of rdg ± 2 digits)

2MΩ

1KΩ

±(1,0% of rdg ± 5 digits)

Range

Resolution

Accuracy

200V

100mV

±(1,2% of rdg ± 10 digits)

600V

1V

±(1,2% of rdg ± 10 digits)

- Input Impendence:  

1MΩ

- Overload protection:  

600V for 200mV range, 

 

effective DC or AC 600V for other ranges

- Frequency Range: 40 to 400Hz

- Response: average, calibrated in rms of sine wave

- over-load protection: 600V effective value

- over-load protection: 600V effective value

6-4 Resistenza

6-3  DC Current

6-2  AC Voltage

Range

Resolution

Function

1mV

Display: read approximate forward 

voltage of diode

Range

Function

Built-in buzzer will sound if resistance is lower 

than 50Ω

6-5 Diode Test

6-6 Continuity 

- over-load Protection: 600V effective value

- forward DC current: approximate 1mA

- Reversed DC voltage: approximate 3.0V

- over-load protection:  

600V effective value

- open circuit voltage:  

approximate 3.0V

Отзывы: