7.5 Recommended Separation Distances between mobile radio equipment and EKUORE PRO®
eKuore Pro® Series must be used in an electromagnetic environment where the radio interferences are controlled. The user can limit the
electromagnetic interferences by maintaining a minimum distance between the mobile radio equipment (transmitter) and eKuore Pro® / eKuore
Pro® amplified as recommended below, according to the maximum power of the radio equipment.
Transmitter’s maximum power
output
W
Separation distances according to the transmitter’s frequency in meters m
150 kHz a 80 MHz d =[(3,5)/
V1] √P
80 MHz a 800 MHz
d =[(3,5)/E1]√P
800 MHz a 1 GHz
d=[7/E1]√P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
The device has no essential performance.
The device may be affected your Wi-Fi connection to radiated electromagnetic fields of frequencies around 2400 MHz and filter change
sound when you apply electrostatic discharge, without affecting all the basic safety equipment.
NOTA 1
: At 80 MHz and at 800 MHz, the highest frequency range is applicable.
NOTA 2
: These guidelines are not necessary applicable in all situations. The electromagnetic propagation is affected by the absorption and
reflections of structures, objects and people.