EKO INSTRUMENTS CO., LTD. - Pyranometer MS-60/60S - Instruction Manual Ver. 4
Pg. 14
2-3. How to Connect MS-60S [See Table 5-4. Wire Color Codes also]
A] How to connect 4-20mA
Connect the output cable end, DC power supply [5 to 36 V], and ammeter as shown in the figure
below. Install a fuse [0.5A] in series between the DC power supply [+] and the MS-60S connection
as shown below for over current protection.
* When precision resistors are connected in parallel and acquired at voltage value, keep the
precision shunt resistor maximum value at 1
50Ω.
B] How to connect 0-1V[optional]
The default output is 4-20mA. If 0 to 1V output is not selected as a purchase option, 0 to 1V
output is off. Or the setting can be changed with the 485 / USB conversion cable and dedicated
software.
Connect the output cable end, DC power supply [5 to 36 V], precision resistor
100Ω [* please
prepare resister separately], and voltage measurement device as shown in the figure below.
Install a fuse [0.5A] in series between the DC power supply [+] and the MS-60S connection as
shown below for over current protection.
Fig. 5-3B. How to connect MS-60S
Power supply
DC5V to 36V
-
+
0 to 1V input
Measurement
Device
+
-
fuse[0.5A]
R
Fig. 5-3A. How to connect MS-60S
Power supply
DC5V to 36V
-
+
4 to 20mA input
Measurement Device
+
-
fuse[0.5A]