Section 5 -
Page 1 of 1
3121-9001-0167_EN.03.00_QT_Manual
04/2016
•
Cleaning may be carried out by wiping the analyzer with a non-abrasive lint free cloth
dampened with a mild detergent or with iso-propyl alcohol wipes.
•
Disinfection of the system can be performed by wiping the analyzer with a non-abrasive
lint free cloth dampened with a non-hypochlorite (bleach) based disinfectant solution such
as Virkon® , in-line with local guidance on prevention of infection.
•
Always power off the analyzer by unplugging the power supply when performing any
cleaning and disinfection operation. Allow all surfaces to fully dry before restoring power.
•
Always power off the analyzer on a weekly basis. Leave off for a minimum of 10 seconds,
then the analyzer may be powered on in order to perform the internal self-checks. To
power off the system, the Quo-Test Analyzer should be in the home screen, then either
switch off at the mains supply or disconnect the power cable from the power socket on
the back of the analyzer.
•
Do not clean or disinfect inside the cartridge test chamber. Contact your local distributor
or customer support for advice.
•
If you spill liquid inside the the cartridge test chamber, call your local distributor or
customer support for advice.
•
If the instrument is stored when not in use, ensure that the main cored is not liable to
damage or kinking.
•
The analyzer is calibrated by the manufacturer and there are no user-serviceable parts.
Note: Virkon® is a trademark of Antec International Limited
Virkon
®
is a trademark of Antec International Limited.
Section 5
Maintenance
Other than restarting the system once a week to allow the system
to perform a self check, no routine maintenance of the Quo-Test
Analyzer other than cleaning is required.