Section 4 -
Page 3 of 8
3121-9001-0167_EN.03.00_QT_Manual
04/2016
Step 5
Lightly touch the pointed tip of the
Blood collector to the top surface of the
droplet of blood, as per the image.
The blood will be taken up by the
collector.
The blood must completely fill the slot.
Note: avoid taking up air bubbles and
excess sample on the outside of the tip
of the Blood collector.
Do not wipe excess blood from the
collector.
Step 6
place the Blood collector into the free
cavity on the top of the Test cartridge,
making sure that the Blood collector is
flush with the top of the Test cartridge
and is not sticking out.
Note: Do not
push the Blood collector
into the Test cartridge, as this may
cause an error message to be displayed
instead of a result.
Refer to the
Troubleshooting Guide and Error
Messages
section of this User manual
for further details.
Running a Test
Step 7
After the Test cartridge is prepared,
open the door of the analyzer and place
the cartridge into the test chamber,
making sure that the Test cartridge is
firmly seated. The Test cartridge must
be inserted into the analyzer within
1 minute of the blood sample being
collected.
5
X
6
7
Содержание Quo-Test
Страница 1: ...User Manual Quo Test Diagnostics for life ...
Страница 32: ...3121 9001 0167 ...