2.2  POWER SUPPLY 

 

Use  ONLY  the power supply packaged with the HemataStat  II. Verify that the 

ON/OFF switch located on the top of the centrifuge is in the OFF " " position.  

Always plug the power supply into the rear of the device first and then plug the three-

prong end into an electrical outlet.  Press the ON/OFF switch to the ON "

ʘ

" position. 

The lid will automatically open and the LCD will display the main menu. 

 

2.3

 

OPTIONAL RECHARGEABLE BATTERY PACK 

 

 

A.  INSTALLATION 

For complete portability, the HemataStat II will operate on a rechargeable 

battery pack, which can be ordered as an option.  The battery pack will need to 

be installed in the device as described below and fully charged before normal 

operation. 

 

 

1.  Make certain the lid is closed and locked and the centrifuge is unplugged 

from the electrical outlet.  Lock the lid by firmly pressing down on the lid 

tab.  

 

2.  Place the device upside down on a smooth, flat surface and on a cloth or 

other protective material to prevent scratching the lid. 

 

3.  Locate the battery pack cover.  Using a small head Phillips screwdriver, 

remove the screws holding the battery pack cover in place. Retain the 

screws.  Remove the battery pack cover. 

4.

 

Connect the battery pack plug into the connector located inside the battery 

pack cavity. Put the connector inside the hole of the housing with the 

connector release tab facing up. Position the battery pack inside the 

compartment.  Lay the cable along side the battery pack to avoid crimping.  

5.  Replace the battery pack cover and secure with the screws provided. 

6.  Plug in the HemataStat II and fully charge the battery pack. 

 

 

B.  CHARGING THE BATTERY PACK 

The battery pack is always charging whenever the HemataStat II is plugged into 

an electrical outlet. Since the battery pack cannot be overcharged, the device 

may remain plugged in continuously without harm. Charge the battery pack 

overnight (or an equivalent period of time).  The percentage of remaining 

battery pack capacity may be determined at any time by performing the 

following procedure. 

Содержание HemataStat II

Страница 1: ...HemataStat II Centrifuge Operator s Manual Separation Technology Inc An EKF Diagnostics Company 570 Monroe Road Suite 1008 Sanford FL 32771 800 777 6668 407 788 3677 www separationtechnology com...

Страница 2: ...4 2 Rotor Loading and Balancing 4 3 Tube Preparation 4 4 Tube Centrifugation 4 5 Reading a Capillary Tube 4 6 Hematocrit Determination Reminders SECTION 5 QUALITY CONTROL AND ASSURANCE 15 5 1 Use of...

Страница 3: ...ages to guide the operator throughout the testing procedure and then displays the test results An optional rechargeable battery pack is available to permit operation in remote locations where alternat...

Страница 4: ...t surface and on a cloth or other protective material to prevent scratching the lid 3 Locate the battery pack cover Using a small head Phillips screwdriver remove the screws holding the battery pack c...

Страница 5: ...here is no longer sufficient capacity to operate the instrument it will fail to complete a cycle and the message CHARGE BATTERY will be displayed 4 If the device is plugged into an electrical outlet w...

Страница 6: ...splay language English Spanish French German or Italian HCT value suffix Choice of displaying sign or only the value itself Decimal point Choice of displaying hematocrit in whole numbers or with a dec...

Страница 7: ...o the main menu D Simultaneously press and release the ENT and RUN buttons The LCD will change to ENT SETUP TUBE I D NEXT RUN E Press RUN to enter the language menu The LCD will display ENT SETUP LANG...

Страница 8: ...ll display ENT SETUP SIGN NEXT RUN I Press ENT to make HCT suffix selection The LCD will display ENT HCT SIGN YES NO RUN J Press ENT again to lock in the suffix or press RUN to remove the suffix When...

Страница 9: ...ONS 4 1 START UP Press the ON OFF switch to the ON position The lid will automatically open and the LCD will display the main menu THE INSIDE DIAMETER ID OF THE TUBES BEING USED MUST MATCH THE SIZE DI...

Страница 10: ...tely or wipe it off with a lint free wipe such as gauze 4 Using your thumb lightly press the finger from above the knuckle to the tip to stimulate blood flow toward the puncture site 5 While applying...

Страница 11: ...op automatically after 60 seconds When the run cycle is completed an automatic braking system will engage to completely stop the rotor An audible tone will indicate that the spin cycle is complete and...

Страница 12: ...the interface Once it has been properly positioned make sure you do not move the tube during the reading process D Press the ENT button The LCD will change to ENT ENTER INTERFACE SEALANT RBCs RUN Mov...

Страница 13: ...ces a line of red blood cells extending from the RBCs PLASMA interface through the PLASMA AIR interface can be observed These fine lines of red blood cells are residuals from the migration and they ha...

Страница 14: ...ayed on the LCD Use only heparinized capillary tubes Spin sample one time 60 seconds ONLY Read the tube within one minute after removing it from the rotor Align vertical black line of the slider at th...

Страница 15: ...r other microhematocrit centrifuges WHEN USING HEMATACHEK Step 1 Ensure bottle cap is tightly closed Step 2 Vigorously tap the bottle against the palm of your free hand As soon as the plastic mixing b...

Страница 16: ...e rpm reading on the LCD should be within 2 of a tachometer reading SECTION 6 MAINTENANCE 6 1 CLEANING As with all electrical devices make sure the centrifuge is unplugged before cleaning Always wear...

Страница 17: ...dically inspect the lid lid gasket and rotor to ensure there are no cracks or damage 6 4 SERVICE To obtain service contact Customer Service at 800 777 6668 407 788 8791 or by email at custserv separat...

Страница 18: ...unit Push gently until the lid opens LOW RPM message Low rpm Insure that the rotor moves freely Check rotor balance RUN ABORTED message Power failure Power supply disconnected Restore power Reconnect...

Страница 19: ...60 109 Unistik 2 Lancets Dual 200 Box 260 110 Unistik 2 Lancets Extra 200 Box 260 111 Unistik 2 Lancets Extra 1 000 Box 270 106 ClearCrit Plastic Microhematocrit Capillary Tubes Heparinized 75mm lengt...

Страница 20: ...d specifically excluding any warranty of merchantability or warranty of fitness for a particular purpose STI will not be liable for any special consequential or incidental damages arising out of the u...

Страница 21: ...rotor stops Front Panel LCD prompts operational modes and alarms Two tactile membrane switch buttons engage various functions and options Speed Designed to operate between 5 670 6 930 rpm and 1 548 2...

Страница 22: ...910 100N 05 2015 Manufactured by Separation Technology Inc 570 Monroe Road Suite 1008 Sanford FL 32771 www separationtechnology com Made in USA...

Отзывы: