Maintenance
Hemo
Point
®
H2
User Manual
49
Illustration 34: Removing the cuvette holder.
The cuvette holder can now be cleaned with a mild soap solution.
For disinfection, standard solvent-free preparations can be used.
To replace the cuvette holder, simply push it in the correct
position into the opening in the housing until it engages.
Please wait until the cuvette holder is completely dry before
replacing it. Do not use any cleaning agent for cleaning the
cuvette holder that could leave scratches on its surface.
6.1.3 Optical
unit
The optical unit is situated inside the analyzer and has no direct
contact with the cuvette therefore no routine cleaning is needed.
Cleaning the optical unit can become necessary if the measured
maximum light intensity of the photometric light source no longer
achieves the appropriate level required for testing.
The optical unit should be cleaned when the following error
message is displayed:
Dirty Optics – Use Optics Cleaner
(see
Illustration 35, Display Screen, Measurement Error).