Sampling and testing procedure
Hemo
Point
®
H2
User Manual
21
Ensure that the cuvette cavity is completely filled and free of air
bubbles.
DO NOT
try and top-off cuvette if not filled completely.
Repeat the process with a new drop of blood and a new cuvette.
Illustration 14: Removing excess blood.
8. In order to avoid contamination of the cuvette holder,
remove excess blood from the outside of the cuvette by
carefully wiping off with a gauze or lint-free tissue.
Please note, do not remove any blood from the cuvette cavity.
DO
NOT
try and top-off cuvette if not filled completely. Repeat the
process with a new drop of blood and a new cuvette.
The cuvette sample can now be tested immediately or within 10
minutes.
DO NOT
read cuvette after 10 minutes. For further
steps, please read Section 4.2 “Performing the test”.
4.1.3 Taking a sample of venous or arterial blood
The
Hemo
Point
®
H2
analyzer can be used for determination of
hemoglobin in venous or arterial blood samples if the blood
collection date is not longer than 24 hours and the sample was
stored refrigerated (2-8°C). Prepare the sample for testing as
follows:
1. Remove sample tube from the refrigerator and bring it to
room temperature (15-30°C).
WARNING!
WARNING!