Product information about
Lactate Scout Check Solution
for use with Lactate Scout Sport
analyzers and Lactate Scout
Test Strips. Please read carefully
before using.
INTENDED USE
Lactate Scout Check Solution are intended to
be used together with the Lactate Scout Sport
analyzers and the Lactate Scout Test Strips to
check the general functionality and accuracy
of the measuring system. No other control
materials are suitable for use with the analyzer.
WARNING NOTICES
Read the user manual of your Lactate Scout
analyzer before using. Do not ingest the Check
Solution. Keep the Check Solution bottle tightly
closed. Keep away from children.
The Check Solution contain a very low
concentration of a toxic substance, which
provides no health risk, avoid ingestion or
contact with mucous membranes. If in doubt
after contact, contact a doctor. The Check
Solution contain buffer solution, additives and
defined lactate concentrations.
STORAGE
Keep the Check Solution protected from light.
Store Lactate Scout Check Solution between
+15°C/59°F and +25°C/77°F. Do not freeze
Lactate Scout Check Solution. Transport and
store the Check Solution only in their original
bottles. Do not fill up, dilute or mix the Check
Solution. After use, close the Check Solution
bottle immediately.
If stored properly, Lactate Scout Check Solution
are stable for 90 days after first opening or until
expiry date, whichever is sooner.
HANDLING
For detailed instructions, please refer to the
“Functionality test” section of the user manual.
(1): Display of the measurement result and
evaluation (check mark or error message)
Ensure that two-digit test strip codes are
entered correctly and match the code on the
test strip vial. Ensure that test strips are stored
and handled correctly. Do not use test strips or
Check Solution that have passed their expiry
date. Never hold the Check Solution bottle
directly to the test strip. This will contaminate
the Check Solution with test strip reagents and
make the remaining Check Solution unusable.
en
Label color:
Red
Yellow
Target value:
5 mmol/L
10 mmol/L
Tolerance range:
4,5 – 5,6 mmol/L
8,9 – 11,1 mmol/L
Lactate Scout
Sport
Functionality test
mode only (1)
Functionality test
mode only (1)
Open the Check Solution bottle and wipe the
opening. Hold the bottle downwards and press
out a small drop. Discard the first drop and press
a second drop onto a clean surface (e.g. a clean
fingernail or a plastic foil). If using the fingernail,
remove residual sweat by washing with water.
Bring the analyzer with the inserted test strip
close to the Check Solution drop until the test
strip tip touches the drop. The required sample
volume (0.2
μ
L) will be drawn into the test
strip automatically. Measurement starts with an
acoustic signal („beep“), and the result will be
displayed. Pull the test strip out of the meter and
dispose of the test strip.
EVALUATION OF THE RESULTS
Use the Check Solution only in the mode
specified for the respective device (see table)
to prevent deviations in results. If the test is
performed correctly, deviations within the
specified tolerance range do not indicate any
malfunctions of the measuring system.
TROUBLE SHOOTING
Should you experience repeated values outside
those defined for the Check Solution and you
have used correctly coded test strips, properly
stored test strips and have followed the user
manuals please contact EKF Customer
Service:
phone +49 39203 511-414
+15°C
+25°C