Lactate Scout

 

Check Solution

To find out more information please  

visit: 

www.lactatescout.info

Vertrieb / Sales 

 

EKF-diagnostic GmbH  

 

Ebendorfer Chaussee 3,  
39179 Barleben, Germany  

phone: 

+49 (0) 39 203 511 414  

email:

 [email protected]

www.ekfdiagnostics.com

Rev.01 01/2022 

SensLab GmbH  

Bautzner Straße 67  
04347 Leipzig, Germany  

phone:

 +49 (0)341 234 1841 

www.senslab.de 

Product information about 

Lactate Scout Check Solution 

for use with Lactate Scout Sport 

analyzers and Lactate Scout 

Test Strips. Please read carefully   

                   before using. 

INTENDED  USE 

 

Lactate Scout Check Solution are intended to 

be used together with the Lactate Scout Sport 

analyzers and the Lactate Scout Test Strips to 

check the general functionality and accuracy 

of the measuring system. No other control 

materials are suitable for use with the analyzer.

WARNING NOTICES 

Read the user manual of your Lactate Scout 

analyzer before using. Do not ingest the Check 

Solution. Keep the Check Solution bottle tightly 

closed. Keep away from children.
The Check Solution contain a very low 

concentration of a toxic substance, which 

provides no health risk, avoid ingestion or 

contact with mucous membranes. If in doubt 

after contact, contact a doctor. The Check 

Solution contain buffer solution, additives and 

defined lactate concentrations.

STORAGE

 

Keep the Check Solution protected from light. 

Store Lactate Scout Check Solution between 

+15°C/59°F and +25°C/77°F. Do not freeze 

Lactate Scout Check Solution. Transport and 

store the Check Solution only in their original 

bottles. Do not fill up, dilute or mix the Check 

Solution. After use, close the Check Solution 

bottle immediately. 

If stored properly, Lactate Scout Check Solution 

are stable for 90 days after first opening or until 

expiry date, whichever is sooner. 

HANDLING 

 

For detailed instructions, please refer to the 

“Functionality test” section of the  user manual.

(1): Display of the measurement result and 

evaluation (check mark or error message)
Ensure that two-digit test strip codes are 

entered correctly and match the code on the 

test strip vial. Ensure that test strips are stored 

and handled correctly. Do not use test strips or 

Check Solution that have passed their expiry 

date. Never hold the Check Solution bottle 

directly to the test strip. This will contaminate 

the Check Solution with test strip reagents and 

make the remaining Check Solution unusable. 

en

Label color: 

Red

Yellow

Target value: 

5 mmol/L

10 mmol/L

Tolerance range:

4,5 – 5,6 mmol/L

8,9 – 11,1 mmol/L

Lactate Scout 

Sport

Functionality  test  

mode only  (1)

Functionality  test  

mode only  (1)

Open the Check Solution bottle and wipe the 

opening. Hold the bottle downwards and press 

out a small drop. Discard the first drop and press 

a second drop onto a clean surface (e.g. a clean 

fingernail or a plastic foil). If using the fingernail, 

remove residual sweat by washing with water. 

Bring the analyzer with the inserted test strip 

close to the Check Solution drop until the test 

strip tip touches the drop. The required sample 

volume (0.2 

μ

L) will be drawn into the test 

strip automatically. Measurement starts with an 

acoustic signal („beep“), and the result will be 

displayed. Pull the test strip out of the meter and 

dispose of the test strip.

EVALUATION OF THE RESULTS 

Use the Check Solution only in the mode 

specified for the respective device (see table) 

to prevent deviations in results. If the test is 

performed correctly, deviations within the 

specified tolerance range do not indicate any 

malfunctions of the measuring system.

TROUBLE SHOOTING 

Should you experience repeated values outside 

those defined for the Check Solution and you 

have used correctly coded test strips, properly 

stored test strips and have followed the user 

manuals  please contact EKF Customer  

Service:

 [email protected],  

phone +49 39203 511-414 

+15°C

+25°C

Содержание Lactate Scout Sport

Страница 1: ...Lactate Scout Sport Instruction manual...

Страница 2: ......

Страница 3: ...written consent of SensLab GmbH Part No 7023 9018 0127 Rev 01 01 2022 Lactate Scout Sport instruction manual Manufacturer SensLab GmbH Bautzner Str 67 04347 Leipzig Germany Sales and service EKF diag...

Страница 4: ...batteries 10 1 4 Switching the device on and off 11 1 5 Display 11 1 6 Code setting for the test strips 12 2 Performing a test 13 2 1 Preparing for a test 13 2 2 Sample collection and test 14 2 3 Chec...

Страница 5: ...aying stored values 26 3 5 Battery level indicator 27 4 Performing a step test 28 4 1 Step test mode 28 4 2 Displaying the post exertion period 30 5 Storage cleaning and disposal 31 6 Error messages a...

Страница 6: ...rips are only for single use Safety instructions PLEASE READ THE INSTRUCTION MANUAL BEFORE USE AND KEEP IT IN A SAFE PLACE Failure to comply with any text marked as a WARNING can result in serious dam...

Страница 7: ...ecified in the instruction manual ATTENTION Make sure you only use accessories that have been provided or recommended by the manufacturer The measuring device must not be used if it is not functioning...

Страница 8: ...menus sub menus and menu items for adjusting values e g time and date OK button 1D for enabling menus sub menus and menu items for confirming choices for activating and deactivating configuration mod...

Страница 9: ...batteries When inserting the new batteries ensure that the polarity is correct Push the battery cover back in until it completely locks into place 4 If the time to replace the batteries exceeds 30 se...

Страница 10: ...sed for two minutes in configuration and display mode or if the device is not used for two minutes while the test strips is inserted in the measuring device 1 5 Display An information line will be dis...

Страница 11: ...n The code setting menu will open The symbol line will display the CODE symbol 7 Enter the two digits of the code starting from the left Enter the digits of the code using the two arrow keys and confi...

Страница 12: ...The display will briefly show the code which is currently configured A blinking drop symbol will then be displayed to indicate that the measuring device is ready to take a measurement The ambient tem...

Страница 13: ...tainer immediately after removing the test strip Never leave the container open 2 2 Sample collection and testing Wash your fingers or earlobes with clean water at the puncture site Dry the puncture s...

Страница 14: ...ustic signal will sound and the test will begin A circular symbol with a progress bar will be displayed 5 After ten seconds a second acoustic signal will sound and the test result will be displayed in...

Страница 15: ...e code is displayed briefly when you switch measuring mode on 7 If more than 24 tests are performed with the same code a reminder to check the code will be displayed for three seconds 8 Pressing the O...

Страница 16: ...nctions and displays Settings can be changed in configuration and display mode In order to go to configuration and display mode there must not be a test strip inserted in the measuring device The conf...

Страница 17: ...on 2C 1B 2 2A 2B 2C 2A 3A 3B 3C 3 3 1 1 Volume control If you choose the symbol 2A in the Device settings menu and confirm your selection by pressing the OK button you will be taken to the volume cont...

Страница 18: ...nd day using one of the two arrow keys and confirm your selection by pressing the OK button Once you have confirmed your selection you will then have the option to set the time A clock will appear in...

Страница 19: ...20 2A 1 2 3 4 5...

Страница 20: ...device firmware and device components information If you use the arrow keys to browse through the pages in this menu item then the following information will be displayed Serial number of the measurin...

Страница 21: ...unctionality test is used to check the correct functioning of the Lactate Scout Sport measuring system If there are any doubts regarding the accuracy of the test result or the correct functioning of t...

Страница 22: ...layed with the measurement unit mmol L in the symbol line The configured code will be displayed briefly after you have inserted the test strip This code must also match the code printed on the test st...

Страница 23: ...You can track the progress of the test on the display 22 and the lactate concentration will be displayed If the result is within the permissible range a tick will be shown in the centre of the display...

Страница 24: ...ice will switch back to the main menu 25 26 1E If you want to stop the stopwatch during the 10 minute interval you can do so by briefly pressing the OK button An acoustic signal will sound several tim...

Страница 25: ...7B The date and time of this measurement will be shown in the bottom section of the display 27C If the measurement was taken as part of a step or endurance test where applicable the display will also...

Страница 26: ...29 There are three different display options available Batteries are full 29A Batteries are partially discharged 29B Batteries are almost fully discharged and batteries need to be changed 29C If the b...

Страница 27: ...port along with the relevant configured parameters For a simple step test protocol see https www ekfdiagnostics com lactate testing for athletes html The general procedure for step test monitoring is...

Страница 28: ...1st phase of the exertion phase and take the sample as described in section 2 6 Remove test strip when result is displayed 7 Repeat steps 5 to 6 during the exertion phase of the step test 5 Note that...

Страница 29: ...the post exertion phase the amount of time since the last test in the exertion phase is also displayed by the measuring device in measuring mode 6 The time display is updated every 10 seconds The post...

Страница 30: ...d battery housing Any sample residue Check Solution dust or other contaminants on the housing must be cleaned using a soft lint free cloth or paper towel To do this moisten the cloth with a mild clean...

Страница 31: ...Error when filling the test strip Sample too small or has run contact time of test strips with sample too short interrupted sample absorption or test strip pressed against skin Repeat the test with a...

Страница 32: ...ns for performing the test Dry your hands well before collecting the sample Use the Lactate Scout Check Solution to perform a functionality test Please contact the Lactate Scout Sport service team if...

Страница 33: ...ctionality test with new materials after checking the code Follow the instructions for performing the functionality test Make sure that the target concentration you have selected in the menu matches t...

Страница 34: ...isplay Time displayed with black background The date and time were not set when the device was configured reset Set the date and time Device fails to turn on Device fails to turn on Device fails to tu...

Страница 35: ...range of the measuring device Change the batteries Please contact the EKF technical support team Defective test results Values seem too high too low For example low values after exertion or high valu...

Страница 36: ...ion Slow update of the screen content Slow update of the screen content The Lactate Scout Sport uses an e paper display Please note that it may take some time for the display to update depending on th...

Страница 37: ...Precision 0 5 6 7 mmol L blood lactate 0 2 mmol L 6 8 25 0 mmol L blood lactate 3 At hematocrit values out of the specified range higher deviations are possible Measuring principle Enzymatic amperome...

Страница 38: ...st allocations Power supply 2 x 3 V CR2450 3V lithium batteries button cell batteries Device dimensions 91 mm L x 46 mm W x 21 mm H Weight 60 g including batteries This product contains the Roboto 201...

Страница 39: ...documentation and on packaging Symbol Meaning Manufacturer Serial number Attention consult accompanying documentation observe safety instructions Biological risks Follow instructions for use Designed...

Страница 40: ...41 Symbol Meaning Test strip number Do not expose to direct solar radiation Protect against liquid Direct current Used batteries must be disposed of in accordance with local laws and regulations...

Страница 41: ...rt measuring device with 2 batteries CR2450 7023 0522 Lactate Scout Test Strips 24 test strips in a vial 7023 3431 Lactate Scout Check Solution Duo Pack 2 levels 1 x 4 5 5 6 mmol L and 1 x 8 9 11 1 mm...

Страница 42: ...e for at least 20 min Before using test strips from deep freezer keep the closed vial at ambient temperature for at least 2 hours During use the Lactate Scout analyzer and test strips should be at the...

Страница 43: ...tions please refer to the Functionality test section of the user manual 1 Display of the measurement result and evaluation check mark or error message Ensure that two digit test strip codes are entere...

Страница 44: ...7023 9018 0127_EN 1 0_LSSport_Manual...

Отзывы: