22
ATTENTION
Never bring the opening of the Check Solution
bottle directly into contact with the test strips.
Doing so will contaminate the contents of the
bottle and render it unusable.
Check Solution, test strips and measuring
device must be at the same temperature when
performing the functionality test.
WARNING
Do not swallow the Check Solution.
Avoid contact with mucous membranes.
Risk of small parts being ingested:
keep the Check Solution out of the reach
of children.
3.2 Functionality test
The functionality test is used to check the
correct functioning of the Lactate Scout
Sport measuring system. If there are any
doubts regarding the accuracy of the test
result or the correct functioning of the
measuring device, you must perform a
functionality test. Please use the Lactate Scout
Check Solution for the functionality test. The
label on the Check Solution bottle states the
range of the Check Solution.
The Check Solution is available in the following
concentrations:
8.9 – 11.1 mmol/L (display: 10 mmol/L)
4.5 – 5.6 mmol/L (display: 5 mmol/L)
The Check Solution can be used for a period
of three months after initial opening. Once
opened, ensure the Check Solution is stored
tightly closed between 15 – 25°C.