background image

Appendix

34

Recommended separation distances between portable or mobile RF communication
equipment and the RadiForce series

The RadiForce series is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the RadiForce series
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the RadiForce series.
Immunity to proximity fields from following RF wireless communication equipments has
been confirmed:

Test

frequenc

y

(MHz)

Bandwid

th

*1

(MHz)

Service

*1

Modulation

*2

Maximu

m power

(W)

Minimum
separatio

n

distance

(m)

IEC /

EN60601 test

level

(V/m)

Complia

nce level

(V/m)

385

380 –
390

TETRA 400

Pulse
modulation

*2

18 Hz

1.8

0.3

27

27

450

430 –
470

GMRS 460,
FRS 460

FM
±5 kHz
deviation
1 kHz sine

2

0.3

28

28

710

704 –
787

LTE Band 13,
17

Pulse
modulation

*2

217 Hz

0.2

0.3

9

9

745

780

810

800 –
960

GSM 800 / 900,
TETRA 800,
iDEN 820
CDMA 850,
LTE Band 5

Pulse
modulation

*2

18 Hz

2

0.3

28

28

870

930

1720

1700 –
1990

GSM 1800;
CDMA 1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25;
UMTS

Pulse
modulation

*2

217 Hz

2

0.3

28

28

1845

1970

2450

2400 –
2570

Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7

Pulse
modulation

*2

217 Hz

2

0.3

28

28

5240

5100 –
5800

WLAN 802.11 a/
n

Pulse
modulation

*2

217 Hz

0.2

0.3

9

9

5500

5785

*1

For some services, only the uplink frequencies are included.

*2

carrier are modulated using a 50 % duty cycle square wave signal.

Содержание RediForce MS236WT-A

Страница 1: ...ad the Instructions for Use and Installation Manual before use Please retain this manual for future reference See the Installation Manual for monitor settings and adjustments The latest product inform...

Страница 2: ...usted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications No part of this manual may be re...

Страница 3: ...es Main power switch Press to turn the monitor s main power off Main power switch Press to turn the monitor s main power on Power button Press to turn the monitor s power on or off Alternating current...

Страница 4: ...PRECAUTIONS 4 Symbol This symbol indicates Mark signifying compliance with UK regulations UK Responsible Person Authorised representative in Switzerland...

Страница 5: ...o service this product yourself as opening or removing covers may result in fire electrical shock or equipment damage WARNING Keep foreign objects or liquids away from the unit Metal parts flammable m...

Страница 6: ...orrosive gases such as sulfur dioxide hydrogen sulfide nitrogen dioxide chlorine ammonia and ozone Do not place in environments with dust components that accelerate corrosion in the atmosphere such as...

Страница 7: ...ury and or equipment damage Before installation make sure that desks walls or any other installation surface has adequate mechanical strength If the unit falls disconnect the power immediately and ask...

Страница 8: ...hen moving the product hold the bottom of the monitor firmly Monitors of size 30 inches and above are heavy When unpacking and or carrying the monitor ensure at least two people are utilized If your d...

Страница 9: ...cleaning it Cleaning the unit while it is plugged into a power outlet may result in electrical shock CAUTION If you plan to leave the unit unused for an extended period of time disconnect the power p...

Страница 10: ...monitor has recovered from power saving mode before making adjustments to it If the monitor displays continuously over a long period of time smudges or burn in may appear To maximize the life of the m...

Страница 11: ...chemicals on a frequent basis Chemicals such as alcohol and disinfectant may cause gloss variation tarnishing and fading of the cabinet or LCD panel surface and also quality deterioration of the imag...

Страница 12: ...chanism 14 1 1 3 10 point Multi touch Support 14 1 2 Package Contents 15 1 2 1 Contents Included in the EIZO LCD Utility Disk 15 1 3 Controls and Functions 16 1 3 1 Front 16 1 3 2 Rear 17 2 Installati...

Страница 13: ...ansport Storage Conditions 26 4 2 Compatible Resolutions 27 4 3 Accessories 27 Appendix 28 Medical Standard 28 Classification of Equipment 28 EMC Information 29 Environments of Intended Use 29 Technic...

Страница 14: ...o touch Stand Mechanism The tilt angle can be adjusted steplessly You can smoothly change the screen position to make it easier to work in a way that suits how you are using the monitor such as for of...

Страница 15: ...EIZO LCD Utility Disk CD ROM Instructions for Use Cleaning cloth x 1 Touch pen x 1 Touch pen holder x 1 Touch pen holder mounting screw x 1 Screw hole cover sheet x 2 Tie band push in type x 1 1 2 1...

Страница 16: ...nals to be displayed 4 button 1 Selects items displayed in the Setting menu adjusts or sets the functions and displays the brightness adjustment screen 5 button 1 Selects items displayed in the Settin...

Страница 17: ...f the angle 7 Tie band mounting hole Attach a tie band to prevent the HDMI cable from disconnecting 8 Stereo mini jack By connecting an external device using a stereo mini jack cable external audio ca...

Страница 18: ...Position the monitor so that there is no light to interfere with the screen 2 2 Installing Place a soft cloth or similar material on a stable desk and open the stand until it makes a clicking sound At...

Страница 19: ...function The stereo mini jack cable is for when you want to get audio from the monitor speakers when the D Sub mini 15 pin cable has been used for connection Stereo mini jack cable Commercially avail...

Страница 20: ...ppear see No Picture Problem 23 for additional advice Attention Depending on the PC settings the display may not show correctly or not be at the appropriate scale Check the settings for the PC if requ...

Страница 21: ...Attention After the adjustment is finished make sure that the cables are correctly connected 2 6 Mounting the Touch Pen Holder Mount the touch pen holder in either of the holes to the left and right o...

Страница 22: ...TPOffset can also be downloaded from our website www eizoglobal com Run TPOffset as necessary in the following cases The usage environment has changed significantly Operability feels changed There is...

Страница 23: ...Turn off the main power and then turn it on again Power switch indicator flashes Orange This symptom may occur when the PC is connected to the DisplayPort connector Use a signal cable recommended by u...

Страница 24: ...eck whether the PC is configured to meet the resolution and vertical scan frequency requirements of the monitor see Compatible Resolutions 27 Reboot the PC Select an appropriate setting using the grap...

Страница 25: ...rosoft Windows 10 32 bit 64 bit Microsoft Windows 8 1 32 bit 64 bit Number of simultaneous touch points Max 10 points 1 EIZO support will end when OS vendor support ends 4 1 3 Video Signals Input Term...

Страница 26: ...60 Hz 0 80 0 45 A Maximum Power Consumption 47 W or less Power Save Mode 0 5 W or less 1 Standby Mode 0 3 W 1 1 No USB upstream port is connected Administrator Settings Compatibility Mode is Off no ex...

Страница 27: ...0 x 600 72 188 800 x 600 75 000 1024 x 768 60 004 1024 x 768 70 069 1024 x 768 75 029 1280 x 720 50 000 1280 x 720 59 940 1280 x 720 60 000 1280 x 960 60 000 1280 x 1024 60 020 1280 x 1024 75 025 1680...

Страница 28: ...ctromagnetic waves that could influence limit or result in malfunction of the monitor Install the equipment in a controlled environment where such effects are avoided Classification of Equipment Type...

Страница 29: ...al precautions regarding EMC and need to be installed You need to carefully read EMC Information and the PRECAUTIONS section in this document and observe the following instructions when installing and...

Страница 30: ...ores Technical Descriptions Electromagnetic emissions The RadiForce series is intended for use in the electromagnetic environment specified below The customer or the user of the RadiForce series shoul...

Страница 31: ...l environment Surges IEC EN61000 4 5 1 kV line to line 2 kV line to ground 1 kV line to line 2 kV line to ground Mains power quality should be that of a typical commercial or hospital environment Volt...

Страница 32: ...e level in each frequency range 2 Interference may occur in the vicinity of equipment marked with the following symbol Note UT is the a c mains voltage prior to application of the test level At 80 MHz...

Страница 33: ...abnormal performance is observed additional measures may be necessary such as reorienting or relocating the RadiForce series 2 Over the frequency range 150 kHz to 80 MHz field strengths should be less...

Страница 34: ...ximu m power W Minimum separatio n distance m IEC EN60601 test level V m Complia nce level V m 385 380 390 TETRA 400 Pulse modulation 2 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviat...

Страница 35: ...150 kHz 80 MHz d 1 2 P 80 MHz 800 MHz d 1 2 P 800 MHz 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power...

Страница 36: ...ful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or telev...

Страница 37: ...the following substances except for the exemptions listed in schedule II Lead Mercury Hexavalent Chromium Polybrominated Biphenyls or Polybrominated Diphenyl Ethers exceeding a concentration of 0 1 b...

Страница 38: ...All rights reserved 1st Edition May 10th 2022 153 Shimokashiwano Hakusan Ishikawa 924 8566 Japan Carl Benz Stra e 3 76761 R lzheim Germany 8 5B 1 Queens Square Ascot Business Park Lyndhurst Road Asco...

Отзывы: