EMC Information
The RadiForce series has a performance that appropriately displays images.
Environments of Intended Use
The RadiForce series is intended to be used in Professional healthcare facility environments such as
clinics and hospitals.
The following environments are not suitable for the RadiForce series to be used:
•
Home healthcare environments
•
In the vicinity of high-frequency surgical equipments such as electrosurgical knives
•
In the vicinity of short-wave therapy equipments
•
RF shielded room of the medical equipment systems for MRI
• In shielded location Special environments
• Installed in vehicles including ambulances.
• Other special environment
WARNING
The RadiForce series requires special precautions regarding EMC and need to be installed. You
need to carefully read EMC Information and the “PRECAUTIONS” section in this document, and
observe the following instructions when installing and operating the product.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent
or stacked use is necessary, the equipment or system should be observed to verify normal
operation in the configuration in which it will be used.
When using a portable RF communication equipment, keep it 30 cm (12 inches) or more away from
any part, including cables, of the RadiForce series. Otherwise, degradation of the performance of
this equipment could result.
Anyone who connects additional equipment to the signal input part or signal output parts,
configuring a medical system, is responsible that the system complies with the requirements of
IEC/EN60601-1-2.
Be sure to use the cables attached to the product, or cables specified by EIZO.
Use of cables other than those specified or provided by EIZO of this equipment could result in
increased electromagnetic emissions or decreased electromagnetic immunity of this equipment
and improper operation.
Cable
EIZO Designated
Cables
Max. Cable Length
Shielding
Ferrite Core
Signal cable (DisplayPort) PP300 / PP200
3 m
Shielded
Without Ferrite Cores
Signal cable (DVI)
DD300DL / DD200DL
3 m
Shielded
With Ferrite Cores
USB cable
UU300 / MD-C93
3 m
Shielded
With Ferrite Cores
Power cord (with earth)
-
3 m
Unshielded
Without Ferrite Cores
26
Appendix