background image

8

Notice for This Monitor

To Use the Monitor for a Long Time

 

Quality control

 •

The display quality of monitors is affected by the quality level of input signals and the degradation of 

the product. Perform visual checks and periodic constancy tests to comply with medical standards / 
guidelines according to your application, and carry out calibration as necessary. Use of the RadiCS 

monitor quality control software enables you to perform high-level quality control that meets medical 

standards / guidelines.

 • It takes about 15 minutes (under our measurement conditions) for the monitor display to stabilize. 

Please wait 15 minutes or more after the monitor power has been turned on or the monitor 

has recovered from the power saving mode before performing various tests for quality control, 

calibration, or screen adjustment of the monitor.

 • We recommend that monitors be set to the recommended level or lower to reduce changes in 

luminosity caused by long-term use and maintain stable brightness.

 • To adjust measurement results of the integrated calibration sensor (Integrated Front Sensor) to 

those of an EIZO external sensor (UX1 or UX2 sensor) that is sold separately, perform correlation 
between the Integrated Front Sensor and the external sensor using RadiCS / RadiCS LE. Periodical 
correlation allows you to maintain the measurement accuracy of the Integrated Front Sensor at a 

level equivalent to that of the external sensor.

Attention

 • The display status of the monitor may change unexpectedly due to an operating error or unexpected setting 

change. Using the monitor with the control buttons locked is recommended after adjusting the screen of the 
monitor. For details on how to set, refer to the Installation Manual (on the CD-ROM).

 

Cleaning

Periodic cleaning is recommended to keep the monitor looking new and to prolong its operation 
lifetime.
Gently wipe off any dirt on the cabinet or panel surface with a soft cloth soaked in a small amount of 
water or one of the chemicals listed below.

Chemicals that may be used for cleaning

Material name

Product name

Ethanol

Ethanol

Isopropyl alcohol

Isopropyl alcohol

Chlorhexidine

Hibitane

Benzalkonium chloride

Welpas

Alkyldiaminoethylglycine

Tego 51

Glutaral

Sterihyde

Attention

 •

Do not use chemicals on a frequent basis. Chemicals such as alcohol and antiseptic solution may cause 
gloss variation, tarnishing, and fading of the cabinet or panel, and also quality deterioration of the image.

 • Never use any thinner, benzene, wax, and abrasive cleaner, which may damage the cabinet or panel.
 • Do not let chemicals come into direct contact with the monitor.

Note

 • It is recommended that ScreenCleaner (available as an option) be used to clean the cabinet and LCD panel 

surface.

To Use the Monitor Comfortably

 • Staring at the monitor for a long time tires your eyes. Take a 10-minute rest every hour.
 • Look at the screen from a proper distance and from a proper angle. 

Содержание RadiForce RX360

Страница 1: ...urself with safe and effective usage Please retain this manual for future reference For monitor adjustment and settings refer to the Installation Manual For the latest product information including th...

Страница 2: ...unit This product has been adjusted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications N...

Страница 3: ...ymbol indicates Main Power Switch Press to turn the monitor s main power off Main Power Switch Press to turn the monitor s main power on Power button Press to turn the monitor s power on or off Altern...

Страница 4: ...not place in any form of transportation ships aircraft trains automobiles etc Do not place in dusty or humid environments Do not place in locations where water may be splashed on the screen bathrooms...

Страница 5: ...and install the unit securely Otherwise the unit may become detached resulting in injury and or equipment damage Before installation make sure that desks walls or any other installation surface has a...

Страница 6: ...or inadequate space Do not use the unit laid down or upside down Blocking the ventilation slots prevents proper airflow and may result in fire electric shock or equipment damage Do not touch the plug...

Страница 7: ...fter a short time period has elapsed depending on the displayed image If this occurs changing the image or leaving the power off for a few hours may solve the problem The backlight of the LCD panel ha...

Страница 8: ...tegrated Front Sensor at a level equivalent to that of the external sensor Attention The display status of the monitor may change unexpectedly due to an operating error or unexpected setting change Us...

Страница 9: ...LCD Utility Disk 11 1 3 Controls and Functions 12 Chapter 2 Installation Connection 13 2 1 Before Installing the Product 13 Installation Requirements 13 2 2 Connecting Cables 14 2 3 Turning On the Po...

Страница 10: ...e PCs Monitor operation from the mouse and keyboard Using the RadiCS RadiCS LE monitor quality control software you can perform the following monitor operations using the mouse and keyboard Switching...

Страница 11: ...and monitor operations For more information about the software or setup procedures refer to RadiCS LE User s Manual Quality control Executing calibration Displaying test results in a list and creating...

Страница 12: ...6 USB upstream port Connect this port to the PC when you use software that needs a USB connection or connect a USB device peripheral device that supports USB to the USB downstream port 7 DVI D connect...

Страница 13: ...ace this product on a lacquer coated desk the color may adhere to the bottom of the stand due to the composition of the rubber Check the desk surface before use Installation Requirements When installi...

Страница 14: ...shapes of the connectors and connect the cables After connecting the DVI cable tighten the fasteners to secure the connector Attention The monitor has two types of DisplayPort connectors input and ou...

Страница 15: ...global com When using a daisy chain connection connect the cable to the input connector To set up a daisy chain connection you need to select Signal Format DisplayPort 1 on the Administrator Settings...

Страница 16: ...t and the PC USB cable UU300 Attention When connecting the monitor to a PC on which RadiCS RadiCS LE has been installed connect the cable to Remove the cap before using Note Switching between PCs that...

Страница 17: ...is recommended that the Power button be turned off When not using the monitor you can turn off the main power supply or disconnect the power plug so that the power is cut completely Note In order to m...

Страница 18: ...rn off the main power and then turn it on again If the signal cable is connected to DisplayPort 1 try switching the DisplayPort version For details refer to the Installation Manual on the CD ROM 2 The...

Страница 19: ...y 2 29 0 Hz 61 5 Hz 720 400 69 Hz 71 Hz Frame Synchronization mode 29 5 Hz 30 5 Hz 59 Hz 61 Hz Dot clock DisplayPort DVI 25 MHz 215 MHz 25 MHz 165 MHz 165 MHz 215 MHz Dual link USB Port Upstream port...

Страница 20: ...ted Resolution H x V Vertical scan frequency Hz DisplayPort DVI Portrait Landscape Portrait Landscape 720 400 70 640 480 60 800 600 60 1024 768 60 1280 1024 60 1600 1200 60 1536 2048 60 1 1 2 1536 204...

Страница 21: ...es that could influence limit or result in malfunction of the monitor Install the equipment in a controlled environment where such effects are avoided Classification of Equipment Type of protection ag...

Страница 22: ...with other equipment If adjacent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will be used When using a portable...

Страница 23: ...lectrostatic discharge ESD IEC EN61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood concrete or ceramic tile If floors are covered wi...

Страница 24: ...site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage...

Страница 25: ...ower W Minimum separation distance m IEC EN60601 test level V m Compliance level V m 385 380 390 TETRA 400 Pulse modulation b 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz...

Страница 26: ...Hz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output po...

Страница 27: ...dio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which...

Страница 28: ...O Corporation All rights reserved http www eizoglobal com Siemensallee 84 76187 Karlsruhe Germany 8 5B 153 Shimokashiwano Hakusan Ishikawa 924 8566 Japan 1st Edition October 2018 Printed in Japan 00N0...

Отзывы: