background image

Notice for This Monitor

Indications for Use

This product is indicated for use in displaying radiological images (including full-field digital 

mammography and digital breast tomosynthesis) for review, analysis, and diagnosis by trained medical 
practitioners.

Attention

 • Mammographic images with lossy compression must not be reviewed for primary image interpretations.

Mammographic images may only be interpreted using an FDA cleared display that meets technical specifications

 

reviewed and accepted by FDA.

 • This product may not be covered by warranty for uses other than those described in this manual.
 •

The specifications noted in this manual are only applicable when the following are used:

 - Power cords provided with the product
 -

Signal cables specified by us

 •

Only use optional products manufactured or specified by us with this product.

Precautions for Use

 • Parts (such as the LCD panel and the fan) may deteriorate over extended periods of time. Periodically 

check that they are operating normally.

 • When the screen image is changed after displaying the same image for extended periods of time, an 

afterimage may appear. Use the screen saver or power save function to avoid displaying the same 

image for extended periods of time. Depending on the image, an afterimage may appear even if it was 

displayed for a short period of time. To remove such a phenomenon, change the image or keep the 

power turned off for several hours.

 • It takes about several minutes for the monitor display to stabilize. Before using the monitor, wait a few 

minutes or longer after the monitor power has been turned on or after the monitor has recovered from 

the power saving mode.

 • If the monitor displays continuously over a long period of time, dark smudges or burn-in may appear. To 

maximize the life of the monitor, we recommend the monitor be turned off periodically.

 •

The backlight of the LCD panel has a fixed lifetime. Depending on the usage pattern, such as usage for 

long continuous periods, the lifespan of the backlight may run out sooner, requiring replacement. When 

the screen becomes dark or begins to flicker, please contact your local EIZO representative.

 • The screen may have defective pixels or a small number of light dots on the screen. This is due to the 

characteristics of the panel itself, and is not a malfunction of the product.

 • Do not press on the LCD panel or edge of the frame strongly, as this may result in display malfunctions, 

such as interference patterns, etc. If pressure is continuously applied to the LCD panel surface, the liquid 

crystal may deteriorate or the LCD panel may be damaged. (If the pressure marks remain on the panel, 

leave the monitor with a black or white screen. The symptom may disappear.)

 • Do not scratch or press on the LCD panel with any sharp objects, at this may result in damage to the 

LCD panel. Do not attempt to brush with tissues as this may scratch the panel.

 • Do not touch the built-in calibration sensor (Integrated Front Sensor). Doing so may reduce the 

measurement accuracy or result in equipment damage.

 • Depending on the environment, the value measured by the built-in illuminance sensor may differ from 

the value shown on a stand-alone illuminometer.

 •

When the monitor is cold and brought into a room or the room temperature goes up quickly, dew 

condensation may occur on the interior and exterior surfaces of the monitor. In that case, do not turn the 

monitor on. Instead wait until the dew condensation disappears, otherwise it may cause some damage 

to the monitor.

English

7

Notice for This Monitor

Содержание RadiForce RX1270

Страница 1: ...yourself with safe and effective usage Please retain this manual for future reference For monitor adjustment and settings refer to the Installation Manual For the latest product information including...

Страница 2: ...ted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications No part of this manual may be repr...

Страница 3: ...tes Main Power Switch Press to turn the monitor s main power off Main Power Switch Press to turn the monitor s main power on Power button Press to turn the monitor s power on or off Alternating curren...

Страница 4: ...ny form of transportation ships aircraft trains automobiles etc Do not place in dusty or humid environments Do not place in locations where water may be splashed on the screen bathrooms kitchens etc D...

Страница 5: ...he unit securely Otherwise the unit may become detached resulting in injury and or equipment damage Before installation make sure that desks walls or any other installation surface has adequate mechan...

Страница 6: ...t install the unit in a place with poor ventilation or inadequate space Do not use the unit laid down or upside down Blocking the ventilation slots prevents proper airflow and may result in fire elect...

Страница 7: ...onitor displays continuously over a long period of time dark smudges or burn in may appear To maximize the life of the monitor we recommend the monitor be turned off periodically The backlight of the...

Страница 8: ...uracy of the Integrated Front Sensor at a level equivalent to that of the external sensor Attention The display status of the monitor may change unexpectedly due to an operating error or unexpected se...

Страница 9: ...Installing the Product 16 Installation Requirements 16 2 2 Connecting Cables 16 2 3 Attaching and Removing RadiLight Focus Spotlight 18 2 4 Turning On the Power 19 2 5 Adjusting the Screen Height and...

Страница 10: ...etween monochrome and color parts of the same image at a pixel level and displays them respectively in optimal gradations High degree of freedom of layout This product has the PbyP Picture by Picture...

Страница 11: ...he monitor allowing users to interpret images effectively even in a dark place RadiLight Focus spotlight RadiLight Focus is the removable spotlight This light illuminates the hands of a user when read...

Страница 12: ...allows you to display or hide the PinP sub window and switch the PC used to operate USB devices at the same time For more information about the setup procedure refer to RadiCS RadiCS LE User s Manual...

Страница 13: ...le reference procedures Readme txt file RadiCS LE monitor quality control software for Windows User s Manual Monitor Installation Manual RadiCS LE User s Manual Outline dimensions RadiCS LE RadiCS LE...

Страница 14: ...ion on how to install and use RadiCS LE refer to RadiCS LE User s Manual on the CD ROM When using RadiCS LE connect the monitor to your PC using the supplied USB cable For more information about how t...

Страница 15: ...in backlight function of the monitor This light illuminates the room indirectly from the back of the monitor 7 Stand The height and angle can be adjusted 8 USB downstream port Connect it to a USB dev...

Страница 16: ...with the screen 2 2 Connecting Cables Attention Check that the monitor PC and peripherals are powered off When replacing the current monitor with this monitor refer to 4 2 Compatible Resolutions page...

Страница 17: ...byP mode to display images from two PCs use of some quality control features such as calibration may be restricted For DisplayPort 1 DisplayPort 2 For One Cable PbyP Signal cable PP300 V14 Signal cabl...

Страница 18: ...SB device USB connected peripheral device to the monitor USB cable UU300 Attention When connecting the monitor to a PC on which RadiCS RadiCS LE has been installed connect the cable to Remove the cap...

Страница 19: ...When not using the monitor you can turn off the main power supply or disconnect the power plug so that the power is cut completely Note In order to maximize the monitor s lifespan by impeding brightn...

Страница 20: ...the Cable Cover 1 Tidy the cables into the cable holder 2 Install the cable cover Example Installing the Cable Cover Right Attention Install the cover properly to ensure ventilation inside the monito...

Страница 21: ...and brightness of LED lights are not consistent even among the products of the same model RadiLight Area may not sufficiently light up the room in the following environment If there is a relatively l...

Страница 22: ...pecified by EIZO Turn off the main power and then turn it on again If the signal cable is connected to DisplayPort 1 try switching the DisplayPort version For details refer to the Installation Manual...

Страница 23: ...kHz 175 kHz HDMI 31 kHz 160 kHz Vertical scan frequency 2 DisplayPort 29 Hz 61 Hz 720 400 69 Hz 71 Hz HDMI 59 Hz 61 Hz 720 400 69 Hz 71 Hz Frame Synchronization mode 29 5 Hz 30 5 Hz 59 Hz 61 Hz Dot c...

Страница 24: ...4 The USB upstream port is not connected DP Power Save is set to On DisplayPort 1 is set to Version 1 2 DisplayPort 2 is set to Version 1 2 and no external load is connected 4 2 Compatible Resolutions...

Страница 25: ...e optional accessories and information about the latest compatible graphics board refer to our web site www eizoglobal com Calibration Kit RadiCS UX2 Ver 5 0 4 or later RadiCS Version Up Kit Ver 5 0 4...

Страница 26: ...that could influence limit or result in malfunction of the monitor Install the equipment in a controlled environment where such effects are avoided Classification of Equipment Type of protection agai...

Страница 27: ...equipment If adjacent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will be used When using a portable RF communic...

Страница 28: ...atic discharge ESD IEC EN61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood concrete or ceramic tile If floors are covered with synth...

Страница 29: ...ould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage prior to applic...

Страница 30: ...inimum separation distance m IEC EN60601 test level V m Compliance level V m 385 380 390 TETRA 400 Pulse modulation b 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0...

Страница 31: ...2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed a...

Страница 32: ...he following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to whi...

Страница 33: ...ZO Corporation All rights reserved 2nd Edition May 7th 2020 Printed in Japan 00N0N165B1 IFU RX1270 6 www eizoglobal com Carl Benz Stra e 3 76761 R lzheim Germany 8 5B 153 Shimokashiwano Hakusan Ishika...

Отзывы: