Technical Descriptions

Electromagnetic emissions

The RadiForce series is intended for use in the electromagnetic environment specified below.

The customer or the user of the RadiForce series should assure that it is used in such an environment.

Emission test

Compliance

Electromagnetic environment - Guidance

RF emissions 

CISPR11 / EN55011

Group 1

The RadiForce series uses RF energy only for its internal function.

Therefore, its RF emission are very low and are not likely to cause any interference in 

nearby electronic equipment.

RF emissions 

CISPR11 / EN55011

Class B

The RadiForce series is suitable for use in all establishments, including domestic 

establishments and those directly connected to the public low-voltage power supply 

network that supplies buildings used for domestic purposes.

Harmonic emissions

 

IEC / EN61000-3-2

Class D 

Voltage fluctuations / 

 

flicker emissions

 

IEC / EN61000-3-3

Complies

Electromagnetic immunity

The RadiForce series has been tested at the following compliance levels according to the testing requirements for 
professional healthcare facility environments defined in IEC / EN60601-1-2.

Customers and users of the RadiForce series must ensure that the RadiForce series is used in the following 
environments:

Immunity test

Test level for 

professional  

healthcare facility  

environments 

Compliance level

Electromagnetic environment - 

Guidance

Electrostatic 

discharge (ESD) 

IEC / EN61000-4-2

±8 kV contact discharge

±15 kV air discharge

±8 kV contact discharge

±15 kV air discharge

Floors should be wood, concrete or ceramic tile. 

If floors are covered with synthetic material, the 

relative humidity should be at least 30%.

Electrical fast 

transients / bursts 

IEC / EN61000-4-4

±2 kV power lines

±1 kV input / output lines

±2 kV power lines

±1 kV input / output lines

Mains power quality should be that of a typical 

commercial or hospital environment.

Surges 

IEC / EN61000-4-5

±1 kV line to line 

±2 kV line to ground

±1 kV line to line 

±2 kV line to ground

Mains power quality should be that of a typical 

commercial or hospital environment.

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines  

IEC / EN61000-4-11

0 % U

(100 % dip in U

T

0.5 cycles and 1 cycle

70 % U

T

 (30 % dip in U

T

25 cycles

0 % U

T

 (100 % dip in U

T

)  

5 sec

0 % U

(100 % dip in U

T

0.5 cycles and 1 cycle

70 % U

T

 (30 % dip in U

T

25 cycles

0 % U

T

 (100 % dip in U

T

)  

5 sec

Mains power quality should be that of a typical 

commercial or hospital environment. If the user 

of the RadiForce series requires continued 

operation during power mains interruptions, it 

is recommended that the RadiForce series be 

powered from an uninterruptible power supply 

or a battery.

Power frequency 

magnetic fields

 

IEC / EN61000-4-8

30 A/m

(50 / 60 Hz)

30 A/m

Power frequency magnetic fields should be at 

levels characteristic of a typical location in a 

typical commercial or hospital environment. The 

product should be kept at least 15 cm away 

from the source of power frequency magnetic 

fields during use.

28

Appendix

Содержание RadiForce RX1270

Страница 1: ...yourself with safe and effective usage Please retain this manual for future reference For monitor adjustment and settings refer to the Installation Manual For the latest product information including...

Страница 2: ...ted specifically for use in the region to which it was originally shipped If operated outside this region the product may not perform as stated in the specifications No part of this manual may be repr...

Страница 3: ...tes Main Power Switch Press to turn the monitor s main power off Main Power Switch Press to turn the monitor s main power on Power button Press to turn the monitor s power on or off Alternating curren...

Страница 4: ...ny form of transportation ships aircraft trains automobiles etc Do not place in dusty or humid environments Do not place in locations where water may be splashed on the screen bathrooms kitchens etc D...

Страница 5: ...he unit securely Otherwise the unit may become detached resulting in injury and or equipment damage Before installation make sure that desks walls or any other installation surface has adequate mechan...

Страница 6: ...t install the unit in a place with poor ventilation or inadequate space Do not use the unit laid down or upside down Blocking the ventilation slots prevents proper airflow and may result in fire elect...

Страница 7: ...onitor displays continuously over a long period of time dark smudges or burn in may appear To maximize the life of the monitor we recommend the monitor be turned off periodically The backlight of the...

Страница 8: ...uracy of the Integrated Front Sensor at a level equivalent to that of the external sensor Attention The display status of the monitor may change unexpectedly due to an operating error or unexpected se...

Страница 9: ...Installing the Product 16 Installation Requirements 16 2 2 Connecting Cables 16 2 3 Attaching and Removing RadiLight Focus Spotlight 18 2 4 Turning On the Power 19 2 5 Adjusting the Screen Height and...

Страница 10: ...etween monochrome and color parts of the same image at a pixel level and displays them respectively in optimal gradations High degree of freedom of layout This product has the PbyP Picture by Picture...

Страница 11: ...he monitor allowing users to interpret images effectively even in a dark place RadiLight Focus spotlight RadiLight Focus is the removable spotlight This light illuminates the hands of a user when read...

Страница 12: ...allows you to display or hide the PinP sub window and switch the PC used to operate USB devices at the same time For more information about the setup procedure refer to RadiCS RadiCS LE User s Manual...

Страница 13: ...le reference procedures Readme txt file RadiCS LE monitor quality control software for Windows User s Manual Monitor Installation Manual RadiCS LE User s Manual Outline dimensions RadiCS LE RadiCS LE...

Страница 14: ...ion on how to install and use RadiCS LE refer to RadiCS LE User s Manual on the CD ROM When using RadiCS LE connect the monitor to your PC using the supplied USB cable For more information about how t...

Страница 15: ...in backlight function of the monitor This light illuminates the room indirectly from the back of the monitor 7 Stand The height and angle can be adjusted 8 USB downstream port Connect it to a USB dev...

Страница 16: ...with the screen 2 2 Connecting Cables Attention Check that the monitor PC and peripherals are powered off When replacing the current monitor with this monitor refer to 4 2 Compatible Resolutions page...

Страница 17: ...byP mode to display images from two PCs use of some quality control features such as calibration may be restricted For DisplayPort 1 DisplayPort 2 For One Cable PbyP Signal cable PP300 V14 Signal cabl...

Страница 18: ...SB device USB connected peripheral device to the monitor USB cable UU300 Attention When connecting the monitor to a PC on which RadiCS RadiCS LE has been installed connect the cable to Remove the cap...

Страница 19: ...When not using the monitor you can turn off the main power supply or disconnect the power plug so that the power is cut completely Note In order to maximize the monitor s lifespan by impeding brightn...

Страница 20: ...the Cable Cover 1 Tidy the cables into the cable holder 2 Install the cable cover Example Installing the Cable Cover Right Attention Install the cover properly to ensure ventilation inside the monito...

Страница 21: ...and brightness of LED lights are not consistent even among the products of the same model RadiLight Area may not sufficiently light up the room in the following environment If there is a relatively l...

Страница 22: ...pecified by EIZO Turn off the main power and then turn it on again If the signal cable is connected to DisplayPort 1 try switching the DisplayPort version For details refer to the Installation Manual...

Страница 23: ...kHz 175 kHz HDMI 31 kHz 160 kHz Vertical scan frequency 2 DisplayPort 29 Hz 61 Hz 720 400 69 Hz 71 Hz HDMI 59 Hz 61 Hz 720 400 69 Hz 71 Hz Frame Synchronization mode 29 5 Hz 30 5 Hz 59 Hz 61 Hz Dot c...

Страница 24: ...4 The USB upstream port is not connected DP Power Save is set to On DisplayPort 1 is set to Version 1 2 DisplayPort 2 is set to Version 1 2 and no external load is connected 4 2 Compatible Resolutions...

Страница 25: ...e optional accessories and information about the latest compatible graphics board refer to our web site www eizoglobal com Calibration Kit RadiCS UX2 Ver 5 0 4 or later RadiCS Version Up Kit Ver 5 0 4...

Страница 26: ...that could influence limit or result in malfunction of the monitor Install the equipment in a controlled environment where such effects are avoided Classification of Equipment Type of protection agai...

Страница 27: ...equipment If adjacent or stacked use is necessary the equipment or system should be observed to verify normal operation in the configuration in which it will be used When using a portable RF communic...

Страница 28: ...atic discharge ESD IEC EN61000 4 2 8 kV contact discharge 15 kV air discharge 8 kV contact discharge 15 kV air discharge Floors should be wood concrete or ceramic tile If floors are covered with synth...

Страница 29: ...ould be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage prior to applic...

Страница 30: ...inimum separation distance m IEC EN60601 test level V m Compliance level V m 385 380 390 TETRA 400 Pulse modulation b 18 Hz 1 8 0 3 27 27 450 430 470 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0...

Страница 31: ...2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 7 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed a...

Страница 32: ...he following measures Reorient or relocate the receiving antenna Increase the separation between the equipment and receiver Connect the equipment into an outlet on a circuit different from that to whi...

Страница 33: ...ZO Corporation All rights reserved 2nd Edition May 7th 2020 Printed in Japan 00N0N165B1 IFU RX1270 6 www eizoglobal com Carl Benz Stra e 3 76761 R lzheim Germany 8 5B 153 Shimokashiwano Hakusan Ishika...

Отзывы: