16
Chapter3Troubleshooting
●
Error code table
Error Code
Description
0***
• ErrorsthatoccurredduringSelfCalibration
1***
• ErrorsthatoccurredduringGrayscaleCheck
*1**
• ErrorsthatoccurredduringDICOM
*2**
• ErrorsthatoccurredduringCAL1
*3**
• ErrorsthatoccurredduringCAL2
**10
• Theproductmaximumbrightnessmaybelowerthanthetargetbrightness.
• Lowerthetargetbrightness.
**11
• Theproductminimumbrightnessmaybehigherthanthetargetbrightness.
• Raisethetargetbrightness.
**34
• Thesensormaynothavecomeoutduringcalibration,orlightmayhaveenteredthe
sensor.
• Turnoffthemainpower,waitforafewminutesbeforeturningthepowerbackon,and
thenexecuteSelfCalibration/GrayscaleCheckagain.
**61
• Thesensormaynothavecomeout.
• Checkwhetherthereisanyforeignobjectnearthesensor.
• ExecuteSelfCalibration/GrayscaleCheckagain.
**95
• TheexecutionconditionsforSelfCalibration/GrayscaleCheckmaynotbeappropriate.
• Checktheexecutionconditionsandresetthemifnecessary.