19
Chapter4ControllingMonitorQuality
4-2. Checking Grayscale
This function allows you to execute Grayscale Check and view the latest result. In addition, when the schedule is
set using RadiCS, the next grayscale check schedule appears.
●
Executing Grayscale Check
Attention
• Do not execute this function when a panel protector (RP-903) is installed.
• Execute Grayscale Check under the actual temperature of the monitor usage environment.
Procedure
1. Choose “RadiCS SelfQC” from the Adjustment menu, and press
.
2. Choose “Grayscale Check” from the “RadiCS SelfQC” menu, and press
.
3. Select “Execute” with or .
4. Press
to exit.
Grayscale Check is executed.
●
Viewing Grayscale Check result
Attention
• The result can be viewed when “QC History” is “On”.
Procedure
1. Choose “RadiCS SelfQC” from the Adjustment menu, and press
.
2. Choose “Grayscale Check” from the “RadiCS SelfQC” menu, and press
.
3. Select “Result” with or .
4. Select “DICOM”, “Custom”, “CAL1” or “CAL2” with or .
5. Press
to exit.
Up to the five latest Grayscale Check results (Passed/Failed/Canceled/ERR) appear.
4-3. Acquiring Life Prediction Data
This function allows you to acquire data necessary to predict the monitor life.
Attention
• When a panel protector (RP-903) is installed, be sure to set “Meas. Frequency” to “Off”.
Note
• Data are collected every 100 hours.
• When you select “Power Save”, data are collected when the mode switches to Power Save or the secondary power off after
the above timing is reached.
• When you select “Routine”, data are saved after the following times have elapsed.
- 500, 1000, 2000, 4000, 7000, 10000, 15000, 20000, 25000, 30000, 35000, 40000, 45000, 50000 hours
Procedure
1. Choose “RadiCS SelfQC” from the Adjustment menu, and press
.
2. Choose “LEA” from the “RadiCS SelfQC” menu, and press
.
3. Select “Meas. Frequency” with or .
4. Select “Power Save”, “Routine”, or “Off” with or .
5. Press
to exit.