27
Appendix
EMC Information
The EX2620-3D has a performance that appropriately displays images.
Environments of Intended Use
The EX2620-3D is intended to be used in Professional healthcare facility environments such as
clinics and hospitals (including use in the vicinity of high-frequency surgical equipment such as
electrosurgical knives).
The following environments are not suitable for the EX2620-3D to be used:
• Home healthcare environments
• In the vicinity of short-wave therapy equipments
• RF shielded room of the medical equipment systems for MRI
• In shielded location Special environments
• Installed in vehicles including ambulances.
• Other special environments
WARNING
The EX2620-3D requires special precautions regarding EMC and need to be installed. You need to
carefully read EMC Information and the "PRECAUTIONS" section in this document, and observe
the following instructions when installing and operating the product.
The EX2620-3D should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation
in the configuration in which it will be used.
When using a portable RF communication equipment, keep it 30 cm (12 inches) or more away from any part,
including cables, of the EX2620-3D. Otherwise, degradation of the performance of this equipment could
result.
Anyone who connects additional equipment to the signal input part or signal output parts,
configuring a medical system, is responsible that the system complies with the requirements of
IEC/EN60601-1-2.
Images may be distorted if the product is used near a device such as high-frequency surgical
equipment. Check in advance that no problems will occur during use.
Be sure to use cables that satisfy the following requirements.
Use of cables that do not satisfy the requirements could result in increased electromagnetic
emissions, reduced electromagnetic immunity of this equipment, and incorrect operations.
Cables
Max. Cable Length
Shielding
AC Mains input
2 m
Un-Shielded
DC Cord
17.5 m
Shielded
BNC Cable (SDI)
30 m
Shielded
DVI Cable
5 m
Shielded
BNC Cable (Composite)
5 m
Shielded
D-Sub Cable
5 m
Shielded
S VIDEO Cable
5 m
Shielded
RS-232C Cable
5 m
Shielded
English