9
HTEC-500
7-1.
Run/Stop
Use the pump key to turn the pump on and off. When the pump is running, the pump light should be on
at the front of the system. Also,
*PUMP*
on the
liquid crystal display (LCD) indicates the pump is running,
while
-PUMP-
indicates the pump is off.
7-2.
Flow Rate Setting
• Select the pump monitor by pushing the display key (the display key switches between the ECD and
the PUMP screen).
• Push the edit key once. ‘ ’ appears beside the flow rate (
μ
l/min).
• Enter the adequate number from 1 to 750 using the 10 number keys.
• Push the edit key once more. ‘ ’ should now be beside pulse free mode.
Fig. 9 Display
*PUMP* Press 0.64MPa
FlowRate
200
μ
L/min
P u l s e F r e e M o d e
O N
P r e s s . L i m i t
1 4 M P a
Indicates the pump is running.
Pressure status
Setting flow rate. (Use the 10 keys for setting)
Indicates the selected setting. Moves and makes changes using the edit key.
ON/OFF using the select key
Use the 10 keys for setting
- E C D -
0 1 . 5 4 n A
A P P. P o t e n
+ 4 0 0 m V
T i m e C o n s t .
1 . 0 s e c
Te m p / S e t 2 5 . 6 / 2 5
˚
C
C
These bars indicate that the ECD is turned off.
Current status
Setting applied voltage (use the 10 keys for setting).
Most sensitive at 1.0 sec.
Noise is reduced at 3.0 sec. Select using the select key.
Indicates the thermocontroller is ON.
Cabinet setting
Actual temperature in the cabinet
Display can be switched between these two screens using the display key.