Homecare Multi-parameter Recorder
- 31 -
250 East Gish Road, San Jose
CA, 95112 USA
Voltage dips, short
interruptions
and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
<5% UT
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of
the
recorder
requires
continued
operation
during
power mains interruptions, it is
recommended that the recorder
be
powered
from
an
uninterruptible power supply or a
battery.
Power
frequency
(50Hz) magnetic field
IEC 61000-4-8
3A/m
3A/m
If image distortion occurs, it may
be necessary to position the
recorder further from sources of
power frequency magnetic fields
or to install magnetic shielding.
The power frequency magnetic
field should be measured in the
intended installation location to
assure that it is sufficiently low.
NOTE UT is the a.c. mains voltage prior to application of the test level.