Homecare Multi-parameter Recorder
- 30 -
250 East Gish Road, San Jose
CA, 95112 USA
Chapter 6 Guidance and Manufacturer’s Declaration of
EMC
Guidance and manufacturer’s declaration – electromagnetic emissions-
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emission
The recorder is intended for use in the electromagnetic environment specified below. The customer or the user
of the recorder should assure that it is used in such and environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
The recorder uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class A
The recorder is suitable for use in all establishments other
than domestic and those directly connected to the public
low-voltage power supply network that supplies building
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Guidance and manufacture’s declaration – electromagnetic immunity
The recorder is intended for use in the electromagnetic environment specified below. The customer or the user
of recorder should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
8 kV air
6 kV contact
8 kV air
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Users must eliminate static in
their hands before use it.
Electrical
fast
transient/burst
IEC 61000-4-4
2 kV for power supply
lines
1 kV for input/output
lines
2kV
for
power
supply lines
1
kV
for
input/output lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Make sure there is not impulse
interference >1kV in use
environment.
Surge
IEC 61000-4-5
1 kV differential mode
2 kV common mode
1 kV differential
mode
2 kV common mode
Mains power quality should be
that of a typical commercial or
hospital environment.