patients; do not use damaged sensor or extension cable. Do not soak the sensor
into water or make it wet, otherwise it may be damaged.
◼
When disposing the disposable SpO
2
probe or useless SpO
2
probe, please
observe all local, state, and federal regulations that relate to the disposal of
this products or similar products.
Caution: In case it is necessary to add a clip to fix the fingertip sensor, the
cable instead of the sensor itself should be clipped. Please note that the cable of
sensor should not be pulled with force.
4.3 Measurement Procedure
1. Selecting SpO2 Sensor
Depending on the patient category, weight and application site, you can select the SpO
2
sensor as required.
2. Connecting SpO2 Sensor
Plug the SpO
2
sensor cable into the SpO
2
connector on the measurement module.
3. Applying SpO2 Sensor
Clean the application site, such as colored nail polish, and apply the sensor
to the patient.
Warning:
◼
Do not use the SpO
2
sensor on a limb where the NIBP cuff is applied. This may
result in inaccurate SpO
2
reading due to blocked blood flow during cuff
inflation.
◼
Do not conduct SpO
2
measurement on the finger smeared with nail polish,
otherwise unreliable measurement results might be produced.
◼
When using finger sensor, make sure the nail faces to the light window.