All-in-1 Operations Manual
Page 40
www.ehealthcareinc.com
Table 1
The All‐in‐1
is intended for use in the electromagnetic environment specified below. The
custom
er or the user of the All‐in‐1Wireless Mobile Monitoring Platform
should assure that it is
used in such an environment.
Emissions test
Compliance
Electromagnetic environment ‐ guidance
RF emissions
CISPR 11
Group1
The All-in-1 Health Monitoring uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The All-in-1 Health Monitoring is suitable for use in all
establishments, including domestic establishments and
those directly connected to the
public low‐voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000‐3‐2
N/A
Voltage fluctuations /
flicker emissions
IEC 61000‐3‐3
Complies