![Edwards T-STATION 85 Скачать руководство пользователя страница 19](http://html.mh-extra.com/html/edwards/t-station-85/t-station-85_instruction-manual_2366171019.webp)
Figure 5 Turbo screen
Start the pump with the turbo screen
When the START/STOP key is pressed, both the turbo and backing pumps start. The display
shows the turbo pump is accelerating by flashing the top left portion of the percentage sign.
When the turbo pump reaches normal speed (>80% default) the percentage sign stops flashing
and remains steady.
If a turbo setpoint (TSP) has been set, only the backing pump starts when the START/STOP key
is pressed. The TSP LED flashes until the setpoint has been reached. Once the setpoint has
been reached the turbo pump starts and the TSP LED turns OFF. If no turbo setpoint has been
set, the TSP LED remains OFF.
Stop the pump with the turbo screen
1. To stop the pumps press the START/STOP key; the display will show "Stop" for three
seconds.
2. Press the ENTER key within that time to stop the pumps. If the ENTER key has not been
pressed, or the START/STOP or NEXT keys are pressed, the display shows the turbo
pump speed and the pumps remain running.
When the turbo pump is decelerating the bottom right portion of the percentage sign
flashes until the pump has completely stopped.
Turbo screen key actions
Table 4 Turbo screen key actions
Keys
Short press
Long press
START
START the turbo, or initialise turbo
STOP sequence, or cancel turbo
STOP, or force turbo STOP when in
error
-
ENTER
Acknowledge turbo STOP
Fault code display when in error
NEXT
Go to gauge screen, or cancel turbo
STOP
Scroll through menu screens
Gauge screen
When the gauge screen is selected, the gauge LED is lit and the gauge pressure reading is
displayed in the selected units.
If no gauge is connected the display shows "---".
D395-94-880A - Operation