Connect to the vacuum system
WARNING:
Connect the turbo pump inlet to the vacuum system before applying power to the T-
Station 85. This will ensure that the pump cannot operate accidentally causing injury.
WARNING:
If installing the vacuum system directly onto the T-Station 85, the weight of the system
must be no more than 10 kg and the centre of gravity must be positioned laterally within
the bounds of the base plate. If this is not the case, the vacuum system must be supported
to ensure that the T-Station 85 does not topple.
1. The turbo pump can be securely fixed to the vacuum system via the inlet flange. Make
sure that the pump inlet and all components fitted to the pump inlet are clean and
dust-free. If the pump inlet is not kept clean, the pump down time may be increased.
2. If the pump has a CF flange, use the copper compression gasket supplied with the
pump and use a full complement of bolts to connect the inlet flange of the pump to
the vacuum system.
3. If the pump has an ISO flange, use a minimum of four claw clamps (each torqued to
10 Nm) to connect the inlet flange of the pump to the vacuum system. Alternatively,
use a rotatable collar and the combined inlet screen and trapped O-ring supplied with
the pump to connect the inlet flange of the pump to the vacuum system; use a full
complement of bolts with the rotatable collar.
4. If the pump has an NW flange, use the centring ring supplied with the pump and a
metal NW clamp to connect the inlet flange of the pump to the vacuum system.
5. All inlet flange bolts must be re-tightened once the system is under vacuum. Ensure
that no torques or other forces are transmitted to the pump from the vacuum system
or the associated pipelines.
Connect to the exhaust extraction system
WARNING:
Vent dangerous gases and gas mixtures safely. Do not expose people to these gases. If
pumping hazardous gases or vapours, observe the safety recommendations of the supplier
of the gas/vapour.
Note:
This section applies to E2M1.5 versions only.
The exhaust system must be configured so that the maximum pressure at the pump outlet
does not exceed 0.5 bar gauge (1.5 bar absolute, 1.5 x 10
5
Pa) at full pump throughput.
1. Make sure that the exhaust pipeline cannot become blocked.
2. If using an exhaust isolation valve, make sure the pump cannot be operated with the
valve closed.