5.8 Shut-down
We recommend, as described in the following procedure, decontaminating the oil
before shutting down the pump; this will prevent damage to the pump by the
contaminates in the oil.
1.
Refer to
on page 38 and decontaminate the oil, as
required.
2.
Close the vacuum system isolation-valve (if not already closed).
3.
Close gas-ballast (set the gas-ballast control to position ‘0’).
4.
On single-phase pumps, use the on/off switch to switch off the pump.
5.
Switch off the electrical supply to the pump.