7. Maintenance and service
7.1 General maintenance
The system requires no user maintenance. Maintenance must be done by trained
personnel. The frequency of the maintenance depends on the process and can be
adjusted according to the user experience. Refer to
on page 39 for the
maintenance schedule.
We recommend a monthly visual inspection of the dry pump system. Check that the
electrical supply cord, hoses and pipelines connected to the pump are in good condition
and are secured properly.
7.2 Maintenance plan
More frequent maintenance may be required if the pump is used to pump aggressive
gases or vapours, solvents, organic substances and acids, or if the pump is operated
continuously at high operating temperature.
Table 15
Maintenance plan
Operation
Frequency
(months)
Service indi-
cator
Reference
Inspect and clean the inlet
strainer
12
No
Inspect and clean the exter-
nal fan cover if required
12
No
on
page 40
Replace the pump bearings
60
Yes
page 40
Replace the pump controller
120
Yes
on page 40
Electrical safety check
60
No
on page
40
Note:
The pump must be returned to the Service Technology Centre for maintenance. For
service indicator codes, refer to
Table: Service indicator codes
7.3 Inspect and clean the inlet strainer
Whenever the pump is disconnected from the vacuum system or annually:
▪
Remove the inlet strainer from the pump inlet, refer to
.
▪
Clean the pump inlet and remove the debris that may have accumulated.
▪
Examine the inlet strainer and if necessary, clean it with a cleaning solution
suitable for the substances pumped.
▪
Install the inlet strainer before you reconnect the pump to the vacuum system.
Refer to
on page 20.
05/2021 - ©Edwards Limited
Page 39
A90301880_C
A90301880_C - Maintenance and service