Operation
A590-00-880 Issue D
4.4
Automatic shut-down
CAUTION
If you select 'Run 'til crash', the pump(s) can be damaged and you may invalidate any warranties on the iH system
equipment.
Normally, if an alarm condition exists, the iH control system will shutdown the iH system. If required you can request
'run 'til crash' operation. In this mode of operation, most alarm conditions will be ignored and the pump(s) will
continue to operate. For safety reasons the following alarms will shut-down the iH system even if you have selected
'run til crash':
Note:
'Run 'til crash' is automatically reset to 'off' when the electrical supply is removed.
z
HCDP CURRENT HIGH/LOW (error numbers 310 and 312).
z
HCDP THERMISTOR HIGH (error number 512).
z
EXHAUST PRESSURE (error number 3912).
On an iH160, 1000 or 1800 system, the following alarms will cause only the HMB pump to shut down, provided 'run til
crash' is not selected:
z
HMB POWER HIGH/LOW (error numbers 810 and 812).
z
HMB PHASE IMBALANCE (error number 1012).
On an iH160, 1000 or 1800 system, the following alarms will cause only the HMB pump to shut down, whether 'run til
crash' is selected or not:
z
HMB CURRENT HIGH LOW (error numbers 710 and 712).
HMB THERMISTOR HIGH (error number 912).
4.5
Unplanned shutdown and alarms
The iH system is fitted with a number of pump protection sensors (refer to
Table 7
) that will give warnings and
alarms. The decision on whether or not to shutdown the pump is left with the tool.
If the iH system has an unplanned shutdown, ensure that the cause of the shutdown is identified and rectified before
restarting. If you are in any doubt, please call a Edwards Service Engineer.
Table 7 - Safety sensors
Safety sensor
Shut-down condition
HCDP pump-motor thermistor
When the thermistor in the HCDP pump-motor indicates that the temperature
of the pump-motor is nominally 150 °C.
HMB pump-motor thermistor
*
*
iH600/1000/1800 systems only.
When the thermistor in the HMB pump-motor indicates that the temperature
of the pump-motor is nominally 150 °C.
Exhaust pressure transducer
†
†
Accessory.
When the pressure in the exhaust pipeline is 9 psig or higher.
Photohelic switch
When the extraction from the exhaust enclosure is lost.