Page 21
W602100880_A - Operation
Figure 6
Measurement signal (Pressure p)
Adjusting the setpoints
The setpoints can be adjusted by the buttons on the gauge.
Adjusting setpoint <1>
1. Push the <SP> button with a pin (maximum ø1.1 mm). The gauge changes to the
switching function mode and outputs the current lower threshold value at the
measurement value output for approximately 10 seconds (LED <1> flashes).
1. Setpoint
2. Hysteresis
3. Threshold value
CAUTION: DAMAGE TO THE EQUIPMENT
Risk of malfunction. If processes are controlled by the signal output, by pushing the <SP>
button, the measurement signal is suppressed and the corresponding threshold value is
output instead. Push the <SP> button only if you are sure that no malfunction will occur.