![Edwards Barocel 7025 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/edwards/barocel-7025/barocel-7025_instruction-manual_2366094014.webp)
Page 12
W602100880_A - Technical data
Analogue measurement signal versus pressure
Figure 2
Analogue measurement signal versus pressure
p = (U
out
/ 10 V) × p (F.S.)
Conversion Torr to Pascal
Torr
mbar
Pa
1.00
1013.25 / 760 = 1.3332…
101325 / 760 = 133.3224…
Example:
Gauge with 10 Torr F.S.
Measurement signal U
out
= 6 V
p
= (6 V / 10 V) × 10 Torr
= 0.6 × 10 Torr = 6 Torr