Figure 2 Dimensions mm (inches)
142 (5.59)
226.5 (8.91)
155 (6.10)
85 (3.34)
11.5 (0.45)
125 (4.92)
110 (4.33)
7
323 (12.71)
155 (6.10)
166.5 (6.55)
75 (2.95)
140 (5.51)
134 (5.27)
19 (0.75)
18 (0.70)
25 (0.98)
30 (1.18)
46.5 (1.83)
54 (2.12)
86.5 (3.4)
20.5 (0.81)
3.4 Lubrication data
Note:
A Material Safety Data Sheet for Ultragrade 15 is available on request.
Table 4
Lubrication data
Recommended oil*
Ultragrade 15
Maximum oil capacity
0.32 litre
* To operate the pump when the ambient temperature is outside the range specified in
Operating and storage conditions on page 12
, or to optimise pump performance when
processing condensible vapours, a different oil may be required.
3.5 Electrical data
The motor start-up current is drawn for less than one second. Use time-lag fuses to
prevent unnecessary fuse failure when the pump starts. If the pump is used at
temperatures lower than 12 °C (53.6 °F), the start-up current will be drawn for longer,
this may cause the motor thermal overload device to open.
E2M0.7 pump
Motor output rating - continuous
0.09 kW
Motor electrical supply
Single phase
E2M1 and E2M1.5 pumps