background image

10 

 

(Off-line experiments are for when you do not have a 
sensor  connected  directly  to  a  computer,  tablet, 
smartphone, or Edu-logger viewer). 
 

Materials needed: 
 

 

900-230 

CO

2

 Sensor 

 

900-215 Battery 

 

Materials needed to configure your offline 
experiment: 
 

 

900-213  Module  or  900-268  module  or  900-272 
Module 

 

USB  to  mini  USB  cable  (included  with  the  900-
213 module). 

 

Procedure: 

 
1.

 

Follow  the  „Sensor  Offset”  guide  in  this 
document. 

2.

 

Connect  the  CO2  sensor  directly  to  the  left  side 
of a charged 900-215 module. 

3.

 

When ready to collect data press the “Start/Stop” 
button on the CO2 sensor, a red light will turn on 
for the duration of the experiment. 

4.

 

To change the experiment settings first connect to 
your  chosen  device  (PC,  Mac,  tablet,  smart 
device)  and  select  the  “Module  setup”  button.  
(For  more  information  on  how  to  change  the 
experiment  settings  view  quick  start  procedure 
section for your chosen device in this document.) 

5.

 

After  the  experiment  has  concluded  (when  the 
“Start/Stop”  button  is  pressed  again  or  the  data 
collection period ends) connect the sensor to your 
chosen device. (full instructions in each device‟s 
quick start procedure section) 

6.

 

Open Edu-logger software. 

7.

 

Click the “Off-line experiment” button. 

Содержание 900-230

Страница 1: ...Edu logger CO2 Logger Sensor Guide 900 230...

Страница 2: ...nical cell in the sensor reacts with carbon dioxide to produce a voltage which is directly interpreted to give a CO2 concentration reading Hundreds of possible experiments that can be done with the 90...

Страница 3: ...Needed 900 230 CO2 sensor One of the following depending on experiment type 900 213 USB Module For On line experiments 900 215 Battery For Off line experiments 900 272 Viewing Module and 900 215 Batt...

Страница 4: ...O2 sensor is detected click the Module setup button Hold the probe vertically so the opening is facing the ground Press and hold the blue Start Stop button the faceplate of the sensor for 3 seconds 3...

Страница 5: ...n connects to a computer via a USB to mini USB cable Please note that you cannot plug the CO2 sensor directly into the computer Resident PC software and browser based application can be downloaded for...

Страница 6: ...on this will open a graph below 9 Click on the Module setup button located on the CO2 sensor icon in the module window to change the sensor settings if need be 10 Click on the experiment set up button...

Страница 7: ...sly connected to the Edu logger network you can run experiments and collect data through a browser of your choosing Procedure 1 Follow the Sensor Offset guide in this document 2 Connect the CO2 sensor...

Страница 8: ...e wifi201 com into the URL bar then wait for 30 60 seconds 11 You will see a Control mode icon in the browser click on this icon 12 The browser will then load a new screen and begin to auto detect the...

Страница 9: ...ger viewer Materials needed 900 230 CO2 Sensor 900 272 Viewing Module 900 215 Battery Procedure 1 Follow the Sensor Offset guide in this document 2 Connect the CO2 sensor to the left side of the viewe...

Страница 10: ...en ready to collect data press the Start Stop button on the CO2 sensor a red light will turn on for the duration of the experiment 4 To change the experiment settings first connect to your chosen devi...

Страница 11: ...1 Connect the 900 213 to a PC Mac tablet or smart device 2 Connect the CO2 sensor to a 900 213 module they directly plug together 3 Open the Edu logger software 4 Click the Tools icon 5 Click the Res...

Страница 12: ...ange and operation modes 350 10 000 ppm ADC resolution 14 bit Resolution 1 ppm Max sample rate S sec 100 Sensor features Fully digital data Rugged plastic ergonomic case Push button switch for Start S...

Страница 13: ...cientific units o C o F Lux ppm for example The sensor is pre calibrated at the factory The built in software in the logger can be upgraded for free at any time using the provided firmware update In s...

Страница 14: ...ion Li2CO3 2Na Na2O 2Li CO2 As a result of the electrochemical reaction electro motive force EMF would be generated according to Nernst s equation EMF Ec R x T 2F ln P CO2 where P CO2 Partial pressure...

Страница 15: ...e CO2 sensor s probe upright with the blue rubber tip pointing downward Do not use the sensor in acid or alkaline solutions with a molarity greater than 1 0 Warranty We promise to deliver our sensor f...

Страница 16: ...16 Thank you for using Edu logger Flexible simple fast forward thinking W www edulab com E info edulab com V10012013...

Отзывы: