EDTM Glass-Chek ELITE GC3200 Скачать руководство пользователя страница 3

Page 3

Navigating Through the Menu

 A small arrow will appear next to the top option.  In the illustration below, the arrow is next to the “MODE” option.  

To move the cursor from line to line, press the UP or DOWN arrow buttons to navigate through the MENU.  To select an option, press the  
POWER button one time.  Once a category is selected, your options will be displayed.  Use the ARROW buttons to scroll through the available 
options for that selection.  Once you have the desired option on the display, press the POWER button one time to select that option.  You will 
then move back one level in the MENU system.  If you make a selection by mistake, press the POWER button again to select that option a 
second time and make your correct selection.  Once you are finished making changes, use the ARROW buttons to move to the EXIT options:    
“EXIT - SAVE CHANGES”   OR    “EXIT - DO NOT SAVE”. 

Selecting Unit of Measure:

The GC3200 can display thickness measurements in various units.  You also have the option of rounding measurements
to the nearest “Glass Thickness Standard”  (Refer to GC3200 Glass Standards chart on page 16).  To choose the display units appropriate for 
your application, follow these steps:

In the MENU, select the “DISPLAY UNITS” option.  The sub-menu to the right will be displayed:

First, select the “DIMENSIONS” option by pressing the POWER button.  Choose if you want your unit of measure to be 

millimeters or inches

.  

Use the ARROW keys to toggle between units and press the POWER button to make your selection.

Next, determine whether you want the meter to:  1.) round measurements to the nearest Glass STANDARD, 2.) display the ACTUAL  glass 
thickness dimensions or 3.) display both STANDARD and ACTUAL.  To turn STANDARD rounding on or off, select the “STANDARD” option.  If 
your dimensions are set to inches, you will have the options:  FRACTION, DECIMAL or OFF.  If your dimensions are set to millimeters, you will 
have the options: ON or OFF.  Use the ARROW button to find your desired setting and press the POWER button to select it.  PLEASE NOTE, if 
the STANDARD option is selected in inches, the glass thickness designations will always be shown in fractions.  If decimal inches are chosen, 
only the air space thickness and overall IG unit thickness will be shown in decimal inches.  The glass thickness will always be shown in 
fractions of an inch when STANDARD is turned on.

Next, choose your ACTUAL setting.  If this setting is activated, your meter will display actual glass thickness dimensions without rounding.  If 
your dimensions are set to inches, you will have the options:  FRACTION, DECIMAL or OFF.  If your dimensions are set to millimeters, you will 
have the options:  ON or OFF.   Use the ARROW buttons to find your desired setting and press the POWER button to select it. 

NOTE:  STANDARD and ACTUAL cannot both be set to OFF.  If they are both activated, the measurements will alternate on the screen after a 
measurement has been taken.  For an illustration of this feature, see the “Measurement Examples” section.

When you exit the Display Units menu, the meter will show an example of what the measurement results screen will look like according to your 
selections.  This will help you determine whether you chose the correct options.  Press the POWER button to exit the example display.
 

Results Screen Display Speed:

Measurement results are displayed across multiple screens.  These screens can be controlled two different ways.  If you take no action, the 
screens will automatically advance from one to the next.  If you do not want to wait for the screen to advance, you can use the ARROW keys to 
advance or go back to a previous screen.  If you let the meter advance the screens for you, the speed at which these screens advance can be 
set in the MENU system.   Select DISPLAY SPEED in the MENU.  You can independently select the number of seconds you want the 
THICKNESS measurement screens to display as well as the LOW-E results screen.  Use the ARROW keys to change the time, and then press 
the POWER button to confirm the selection.  Exit when complete to go back to the main MENU.  PLEASE NOTE:  If you would prefer that the 
meter only advances to the next result screen when you push the ARROW key, then change the Result Screen times to the maximum value of 
60 seconds for both the THICKNESS and LOW E screens.  This will maximize the time each screen is displayed, allowing you to advance the 
screens via the ARROW buttons when ready (as long as you do so before 60 seconds).                                 

 

Содержание Glass-Chek ELITE GC3200

Страница 1: ...n MENU Options Page 3 Navigating the MENU Unit of Measure Results Screen Speed Page 4 Operating MODES Page 5 Language Selection Back light Settings Stored Data Page 6 Proper Operation BEST Location fo...

Страница 2: ...G MENU screen appears Release the ARROW button once that screen appears You can enter the MENU system any time the meter is powered on by pressing and holding one of the ARROW buttons Please note that...

Страница 3: ...in fractions If decimal inches are chosen only the air space thickness and overall IG unit thickness will be shown in decimal inches The glass thickness will always be shown in fractions of an inch wh...

Страница 4: ...t results If you need to know the air space of the IG change the MODE to NORMAL GLASS NO LAMI and the GC3200 will measure the IG as a regular double pane window including the air space See the applica...

Страница 5: ...ff the back light until the measurement results are obtained This helps to extend the life of the battery Stored Data The GC3200 will automatically save the most recent 30 measurements The stored data...

Страница 6: ...st windows in free air not resting on a tabletop surface or stacked on other pieces of glass BOTTOM CENTER Above Back of GC3200 meter with the two lenses circled Use clean compressed air to clean the...

Страница 7: ...the meter did not detect a low E coating on G1 or G2 Since the GC3200 only detects coatings on the first 2 panes you may be required to test both sides of a triple pane window to verify the presence...

Страница 8: ...by showing the results as a combination of G1 L1 G2 When this occurs the meter was not able to cleanly identify the laminate inner layer surfaces Screen 4 below shows an example of how this will appe...

Страница 9: ...clear of any Low E coating In this example the clear glass lami 162 or 5 32 On screens 2 and 4 you will see that these two thickness values are denoted by the letter D D1 D2 Because the meter does no...

Страница 10: ...itional screen to the rotation of results screens The additional screen is called the ASSUMPTION SCREEN If the user knows that both pieces of glass in the laminated piece that are referenced in the AS...

Страница 11: ...to showing a D dimension that groups at least one component into the dimension The D dimension could include a variety of component thickness values in its result For instance the D dimension could i...

Страница 12: ...than one low e coating in the window the Glass Chek ELITE will only identify the first nearest Low E coating that it encounters Testing from the opposite side of the window would allow you to identify...

Страница 13: ...for low E measurement See the Specifications section on Page 15 of this manual for more details on the GC3200 s maximum thickness ranges Battery and replacement The GC3200 meter is powered by a 9 vol...

Страница 14: ...takes varying amounts of time depending on the application Be sure to hold the meter steady against the window until the results are displayed Measurements that will take longer include Laminated Gla...

Страница 15: ...ic Glass The GC3200 will display the thickness of single pane monolithic glass up to 2 75 inches or 70mm thick GLASS 1 GLASS 1 SINGLE LITE GLASS GLASS 2 GLASS 2 AIR SPACE 1 AIR SPACE 1 AIR SPACE 2 GLA...

Страница 16: ...assumes no liability for the consequential damages of any kind through the use or misuse of the GC3200 product by the purchaser or others No other obligations or liabilities are expressed or implied...

Отзывы: