Magg™ 310
Document: umed018307
Revision: A01
Created by: ST
Date: 24 Feb 2021
IPN: 3097222-A01
Source Location: C:\epdm\isleng\products\ed-magghd\manuals\umed018307.docm
Page 17 of 30
User Ma
nual
SubConn Connector: Lubrication and Cleaning
•
Periodically apply Molykote 111 silicone grease or equivalent before mating connectors.
•
For dry mate connections, a layer of grease corresponding to 1/10 the socket depth should
be applied to the female connector.
•
After greasing, fully mate the male and female connector and remove excess grease from the
connector joint.
•
General cleaning and removal of sand or mud on a connector should be performed using a
spray-based contact cleaner like isopropyl alcohol.
Impulse Connector: Lubrication and Cleaning
•
Lubricate mating surfaces regularly with 3M Silicone spray or equivalent, DO NOT GREASE
•
Lubricate O-rings with Molykote 111 or equivalent.
•
Use dust caps to protect connectors wherever possible.
•
Clean connectors with soap and fresh water, rinse out with alcohol and allow connector to air
dry before using.