SE-1 Electrocardiograph User Manual Cleaning, Care and Maintenance
- 42 -
Mild near neutral detergent
Ethanol (75%)
Isopropanol (70%)
Cleaning agents should be applied or removed using a clean, soft, non-abrasive cloth or paper
towel.
11.2.1 Cleaning the Main Unit
WARNING
Turn off the power before cleaning. The mains supply must be switched off if it is used.
1.
Switch off the main unit and disconnect it from the power cord.
2.
Wipe the exterior surface of the equipment using a soft cloth dampened with the cleaning
solution until no visible contaminants remain.
3.
Wipe off the cleaning solution with a fresh cloth or towel dampened with tap water after
cleaning until no visible cleaning agent remains.
4.
Dry the main unit in a ventilated and cool place.
11.2.2 Cleaning the Patient Cable
1.
Wipe the patient cable with a soft cloth dampened with the cleaning solution until no visible
contaminants remain.
2.
Wipe off the cleaning solution with a fresh cloth or towel dampened with tap water after
cleaning until no visible cleaning agent remains.
3.
Wipe off with a dry cloth to remove residual moisture.
4.
Leave the patient cable to air dry.
CAUTION
Any remainder of cleaning solution should be removed from the main unit and the patient
cable after cleaning.
11.2.3 Cleaning the Reusable Electrodes
1.
Wipe off with a soft cloth to remove residual gel.
2.
Wipe the suction bulbs of chest electrodes and the clamps of limb electrodes with a soft cloth
Содержание SE-1
Страница 2: ......
Страница 7: ...V Appendix 2 EMC Information 56 Appendix 3 Abbreviation 60 ...
Страница 68: ......