Vital Signs Monitor User Manual Intended Use and Safety Guidance
- 6 -
CAUTION
12
A ventilated environment is required for monitor installation. Do not block up the
ventilation grille at the back of the device.
13
The device must be connected to the ground to avoid signal interference.
14
To protect eyes from damage, don't look directly at supplementary light for long time.
15
Poor connection might be caused by frequently plugging and unplugging the power
cord. Check the power cord regularly and replace it in time.
NOTE:
1
Position the device in a location where the operator can easily see the screen and
access the operating controls.
2
The monitor can only be used on one patient at a time.
3
If the monitor gets damp or liquid pours on the monitor, please contact the service
personnel of EDAN.
4
This monitor is not a device for treatment purposes.
5
The pictures and interfaces in this manual are for reference only.
6
Regular preventive maintenance should be carried out every two years. You are
responsible for any requirements specific to your country.
7
When there's measurement beyond range, invalid measurement or no measurement
value, it will display -?-.
8
In normal use, the operator shall stand in front of the monitor.
1.3 Explanation of Symbols on the Monitor
1
DEFIBRILLATION-PROOF TYPE BF APPLIED PART
2
TYPE BF APPLIED PART
3
Caution
4
MR Unsafe - Keep away from magnetic resonance
imaging (MRI) equipment
5
Equipotentiality
6
Power Supply switch