Appendix 2 Signal Input/Output Connector
Accessory equipment connected to these interfaces must be certified according to the respective
IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for
medical equipment). Furthermore all configurations shall comply with the valid version of the
system standard IEC/EN 60601-1-1. Everybody who connects additional equipment to the signal
input connector or signal output connector configures a medical system, and is therefore
responsible that the system complies with the requirements of the valid version of the system
standard IEC/EN 60601-1-1. If in doubt, consult our technical service department or your local
distributor.
RJ45/DB9 Interface
(1) RJ45 Interface
RJ45 interface is 1500V AC isolated intensity and the maximum voltage applied should not
15V DC.
(2) DB9 Interface
Pin
Signal
Input/Output
1
TX
Output
2
RX
Input
3
0V Ref.
4
TA
Output
5
TB
Output
6
RA
Input
7
RB
Input
8
485EN
Input
Pin
Signal
Input/Output
1
+5V
Output
2
RX
Input
3
TX
Output
4
485EN
Input
5
0V Ref.
6
TA
Output
7
TB
Output
8
RA
Input
9
RB
Input