Cadence Fetal Monitor User Manual
- 20 -
Removing Paper Jam
When the recorder functions or sounds improperly, open the recorder door to check for a paper
jam. Removing the paper jam in the following way:
Z
Cut the record paper from the paper notch edge.
Z
Open the door of recorder, and revolve the left gear of the recorder.
Z
Pull the paper from below.
Z
Reload the paper.
4.7
Power on the Monitor
WARNING
: If any sign of damage is detected, or the monitor displays some error
messages, do not use it on any patient. Contact biomedical engineer in
the hospital or our service engineer immediately.
Turn the power switch ON, you can hear the short sound of
“
Do
”
, and see that the power ON
indicator is lit. The numeral LED will flash in red color and green color to test it. At the same
time, the system will print a test baseline. About ten seconds or so, the system will enter
monitoring screen after self-test, and you can perform normal monitoring now.
After self-test, the numeral LED will display as Figure 4.3 (dual configuration).
Figure 4.3 Numeral LED display status after self-test (external TOCO)
4.8 Connect Transducers and Cables
Prior to every time use, check for visible damages of the transducers and cables. If damage is
found, replace them at once.
Connect all the necessary transducers, and cables between the monitor and the patient.
Note
:
1) Only connect the transducers and cables supplied by EDAN to the monitor.
2) Place the arrow symbol on the transducer connector on top when connecting
them to the monitor.
3) Be gentle when connecting and using transducers to avoid mechanical damage
to them.