11.2. Flow measurement (option)
This measurement can be done with a
pitot tube. At first the pitot factor of the
pitot
tube
must
be
entered
(
„Adjustments“
/
„Internal“
/
„Pitot
factor“
). After connecting the pitot tube
to the instrument, the zero point of the
sensor can be set with
<F4>
. With
<F1>
the cross section of the flow channel
can be entered (needed for calculation
of the flow rate). After the pitot tube is
positioned in the flow channel, the
display shows the speed (m/s), the flow
rate (Nm
3
/h) and the differential
pressure (Pa). If the value indicated is
stable, press
<Record>
to store the
value in the intermediate memory. If a
printout of the values should occur
simultaneously to the recording in the
intermediate memory, press
<F2>
(the
complete content of the intermediate
memory will be printed out).
Connections for
pitot tube
Measurement stored in
intermediate memory
V.Gas
0.4
m/sec
M.Flow
44
Nm3/h
dP
0.1
Pa
Flow measurem.
25.11.17
recorded!