Page 52
ecom-
J2KN
pro
TECH
17. Technical data
Parameter
Range
Principle
O
2
0 ... 21 vol.%
Electrochemistry
CO (option)
0 ... 1000 ppm
Infrared
CO (option)
0 ... 2500 ppm
Electrochemistry
CO% (option)
2500 ... 63000 ppm
Electrochemistry
NO (option)
0 ... 1000 ppm
Chemiluminescence
NO (option)
0 ... 5000 ppm
Electrochemistry
NO
2
(option)
0 ... 1000 ppm
Photoacoustic
NO
2
(option)
0 ... 1000 ppm
Electrochemistry
NO
2
/SO
2
(option) 0 ... 100 (2000) ppm
Ultraviolet
SO
2
(option)
0 ... 1000 ppm
Infrared
SO
2
(option)
0 ... 5000 ppm
Electrochemistry
H
2
S (option)
0 ... 1000 ppm
Electrochemistry
H
2
(option)
0 ... 2000 ppm
Electrochemistry
C
x
H
y
(option)
0 ... 4 vol. % (CH
4
)
Catalytic
C
x
H
y
(option)
0 ... 2000 ppm (C
3
H
8
)
Infrared
C
x
H
y
(option)
0 ... 30000 ppm (CH
4
)
Infrared
CO% (option)
0 ... 63000 ppm
Infrared
CO
2
(option)
0 ... 20 vol. %
Infrared
CO
2
0 ... CO
2max
Calculation
Air pressure
300 ... 1100 hPa
DMS bridge
T-Gas
0 ... 500 °C
NiCr/Ni
T-Air
0 ... 99 °C
Semi-conductor
Differential pressure
0 ... +/- 100 hPa
DMS bridge
Efficiency
0 ... 120 %
Calculation
Losses
0 ... 99,9 %
Calculation
Excess air
1 ...
∞
Calculation
Ref. O
2
adjustable
0 ... 21 vol.%
Calculation
Taupunkt der Abgase
Calculation
Power supply
Mains power 230 V / 50 Hz~
Battery 7,2 V / 11,6 Ah
Dim. (L x H x D)
525 mm x 845 mm x 270 mm
Weight
approx. 34 kg with heated sampling system
Application limits
+5 °C to +40 °C;
max. 90 % RH, non-condensing
Subject to technical changes
V3.83 / 06.2018
Содержание J2KNpro Tech
Страница 1: ...Operating Instructions ...