The PTC 400 is designed to be a versatile, easy to use temperature calibrator with a simple
user interface. The following instructions will allow the user to begin simple calibration tasks by 
learning the basic operation of the keys and their functions.

2. Quick Start Instructions

A. Key Functions

Key     

    

Function

 
Input/Output    

Toggles the function selected from measurement mode to source mode.

TC

       

 

 

Pressing the “TC” or “Thermocouple” key toggles through all available TC     

          

 

 

types as well as direct mV output.

RTD

     

 

 

Pressing the RTD key toggles through all available RTD types as well as  

  

          

 

 

direct ohms output.

°F/°C

    

 

 

Toggles between °F and °C

0-9 keys

    

 

These keys allow a user to enter an output value.

          

 

 

Example: to output 20 mV select mV output and Press the “2” key then  

  

          

 

 

the “0” key followed by the “ENTER” key.

Arrow

   



   

These keys allow small changes to be made to an output value. Press  

  

          

 

 

either the up or the down key to set the value as desired.

CE       

 

 

The clear entry key allows the user to clear a value before the enter key is     

          

 

 

pressed.

 

Shift

    

 

 

This key has a blue text color and acts as a second function key to all  

  

          

 

 

keys that have an associated second function.

B. Setpoints

 

1.   Storing Setpoints

   The setpoint mode allows up to 9 setpoints to be set in non-volitile memory for each range.     
   The procedure is as follows:
   a. Set the output to the desired value.
   b. Press shift followed by the “SET” key
   c. Choose the desired setpoint storage location by pressing a key (1-9).
   d. Enter a new value and repeat steps 1 through 3 as needed, up to 9 setpoints for each  

  

       range can be stored.

   2. Recalling Setpoints

   a. To recall a setpoint press shift followed by the”SPT” key.
   b. You will be prompted to enter the number of the desired setpoint location set previously.
   c. Press the desired location number and the unit will go to that output.

Содержание PTC 400

Страница 1: ...Bedienungsanleitung Instruction Manual Manuel d Instructions PTC 400...

Страница 2: ...iagrams Instructions 12 A Sourcing Resistance 12 B Sourcing a Thermocouple Signal 12 C Measuring Temperature Using a Thermocouple 12 D Measuring Resistance RTDs 12 E Sourcing Resistance RTDs 13 4 Main...

Страница 3: ...ey Arrow These keys allow small changes to be made to an output value Press either the up or the down key to set the value as desired CE The clear entry key allows the user to clear a value before the...

Страница 4: ...nter 3 for this question c The next screen prompts you to enter the dwell time from 5 to 500 seconds Simply enter a number in that range d Auto step will begin e To exit auto step press the CE key or...

Страница 5: ...tion Figure 2 Note For best accuracy allow a 10 minute warm up period after the PTC 400 is turned on C Measuring Temperature Using a Thermocouple Figure 3 D Measuring Resistance RTDs When measuring re...

Страница 6: ...des Measuring a 2 Wire RTD or Using the Continuity Function Measuring a 3 Wire RTD Measuring a 4 Wire RTD Figure 4 E Sourcing Resistance Into a RTD Transmitter 1 As described previously in section A t...

Страница 7: ...400 that hold the enclosure together NOTE Two of the screws are under the battery compartment door 2 With the case top open locate the two socketed mini fuses mounted near the input output connection...

Страница 8: ...ded Range in C Min Max Accuracy J Thermocouple 200 0 0 0 0 6 C 0 0 1200 0 0 4 C K Thermocouple 200 0 0 0 0 8 C 0 0 1370 0 0 5 C T Thermocouple 200 0 0 0 0 8 C 0 0 400 0 0 5 C E Thermocouple 200 0 0 0...

Страница 9: ...ered neglected damaged by accident or subjected to abnormal operating conditions or improper handling In the event of a warranty claim the faulty device should be sent in We reserve the right to read...

Страница 10: ...ts GmbH Industriestr 2 D 97959 Assamstadt Tel 49 0 62 94 42 24 0 Fax 49 0 62 94 42 24 90 E Mail sales ecom ex com Internet www ecom ex com 5400 AL 01 A00 10 05 nderungen vorbehalten Subject to change...

Отзывы: