4. Maintenance

A. Power Requirements

The PTC 400 operates on 4 AA alkaline batteries or optional rechargeable Ni-Cd batteries. To
replace the alkaline batteries or to install rechargeable batteries remove the two (2) screws on 
the rear battery door and lift the battery cover. When using Ni-Cd batteries the AC adapter will
simultaneously power the PTC 400 and charge the batteries. A full charge will take 10- 12 hours. 
The AC adapter can be used to power the unit on the bench with or without Ni-Cd batteries 
installed. Never connect the AC adapter with alkaline batteries installed, be sure to remove the 
alkaline batteries before using the AC adapter.

B. Field Replacement Fuse Update

PTC 400s come with a field replaceable minifuse. This fuse protects the PTC 400 from mis-
connections beyond the ratings specified for the PTC 400 inputs and outputs. If an overload 
condition occurs and a blown fuse is suspected perform the following fuse check/replacement 
procedure:
1. Remove the four screws on the rear of the PTC 400 that hold the enclosure together.
  NOTE: Two of the screws are under the battery compartment door.
2. With the case top open, locate the two socketed mini fuses mounted near the input/output
  connection jacks.
3. Remove one fuse at a time and check continuity using an ohmmeter. A good fuse will read     
10 ohms or less. A blown fuse will have a very high reading and generally show as an open     
circuit.
4. If a blown fuse is found replace with the enclosed spare fuse.
5. To order more fuses contact ecom instruments GmbH.

C. Calibration

The PTC 400 should hold its rated specifications for a minimum of one year. Given this, annual 
re-calibration is required for best performance. 

5. Specifications

T/C Ranges:

J, K, T, E, R, S, N, B, L, U including – 10 to 75 mV range

RTD Ranges:

Pt 385 (100, 200, 500, 1000 ohms)
Pt 392, JIS, Ni 120, CU10, YSI 400

Ohms Ranges:

0 to 400.00 and 400.0 to 3200.0

Accuracy: 

T/C Type J: ±0.3°C ±10 µV (±0.4°C total error)
RTD PT100: ±0.1°C ±0.075 Ω (±0.3°C total error)
Ohms (400): ±0.1 Ω
Ohms (3200): ±1.0 Ω

RTD IEX-range:

0.01 mA to 3 mA

RTD Frequency Response:

10 mS – works with all pulsed XMTR’s

Oper. Temp: 

-10°C to 50°C

Storage Temp:

-40°C to 60°C

Power Supply: 

4 AA Alkaline Cells, 30 hours typical operating time

Size: 

7.5“ x 4“ x 1.5“ (190 x 102 x 38 mm)

Содержание PTC 400

Страница 1: ...Bedienungsanleitung Instruction Manual Manuel d Instructions PTC 400...

Страница 2: ...iagrams Instructions 12 A Sourcing Resistance 12 B Sourcing a Thermocouple Signal 12 C Measuring Temperature Using a Thermocouple 12 D Measuring Resistance RTDs 12 E Sourcing Resistance RTDs 13 4 Main...

Страница 3: ...ey Arrow These keys allow small changes to be made to an output value Press either the up or the down key to set the value as desired CE The clear entry key allows the user to clear a value before the...

Страница 4: ...nter 3 for this question c The next screen prompts you to enter the dwell time from 5 to 500 seconds Simply enter a number in that range d Auto step will begin e To exit auto step press the CE key or...

Страница 5: ...tion Figure 2 Note For best accuracy allow a 10 minute warm up period after the PTC 400 is turned on C Measuring Temperature Using a Thermocouple Figure 3 D Measuring Resistance RTDs When measuring re...

Страница 6: ...des Measuring a 2 Wire RTD or Using the Continuity Function Measuring a 3 Wire RTD Measuring a 4 Wire RTD Figure 4 E Sourcing Resistance Into a RTD Transmitter 1 As described previously in section A t...

Страница 7: ...400 that hold the enclosure together NOTE Two of the screws are under the battery compartment door 2 With the case top open locate the two socketed mini fuses mounted near the input output connection...

Страница 8: ...ded Range in C Min Max Accuracy J Thermocouple 200 0 0 0 0 6 C 0 0 1200 0 0 4 C K Thermocouple 200 0 0 0 0 8 C 0 0 1370 0 0 5 C T Thermocouple 200 0 0 0 0 8 C 0 0 400 0 0 5 C E Thermocouple 200 0 0 0...

Страница 9: ...ered neglected damaged by accident or subjected to abnormal operating conditions or improper handling In the event of a warranty claim the faulty device should be sent in We reserve the right to read...

Страница 10: ...ts GmbH Industriestr 2 D 97959 Assamstadt Tel 49 0 62 94 42 24 0 Fax 49 0 62 94 42 24 90 E Mail sales ecom ex com Internet www ecom ex com 5400 AL 01 A00 10 05 nderungen vorbehalten Subject to change...

Отзывы: