27
- In potentially explosive areas, the device must be operated only when in its leather
pouch. Please ensure that the leather pouch
is always closed completely and is undamaged while using.
- Only the accessories approved by ecom instruments GmbH must be used.
- Avoid using the device in corrosive acidic or alkaline mediums.
- Please ensure that the device is not taken into Zone 0 areas.
- The supplied thermocouple must be used only outside the Ex-area.
4.1 Special conditions for use in Zone 0 areas
Temperature measurement using the Ex-MX4 BT is allowed within Zone 0 only if the
following conditions are met:
1. The Ex-MX4 BT temperature measuring device must not be brought into Zone 0.
2. Absolutely ensure that the device does not fall into Zone 0, even by mistake. To
ensure this, the device‘s carrying sling must also be attached securely to the
operator’s belt or to a belt loop. The length of the carrying sling should be measu-
red such that the Ex-MX4 BT is still at a distance of 50 cm from the boundary area
of Zone 0 when it hangs freely on the carrying sling.
Only carrying slings approved by ecom instruments GmbH must be used.
5. Safety instructions
- Caution: laser beam!
- Do not look directly into the laser beam!
- Do not aim the beam at other people!
- Danger of eye damage!
- Laser class 2
- Power < 1 mW
- Wave length 635 nm
6. Ex data
EC-type examination certificate no.: TÜV 00 ATEX 1597 X
Ex designation:
II 2 G EEx ia IIC T4
Approved for Zones 1 and 2, device group II, gas group C potentially explosive gases,
vapours or mist, temperature class T4.
7. Technical specifications
Measuring range:
-30°C ... +900°C
Display resolution:
0.1°C
Ambient temperature:
0°C... +50°C;
laser operation up to a maximum of 45°C
Storage temperature:
-20°C ... +50°C (without batteries)
Batteries / power supply:
2 x LR06 (AA) type approved in compliance with IEC