11
21.
Turn the system on.
a.
If you have the control box turn that on as well
.
22.
If you have the control box and level sensor, then jump to the “calibrating the level sensor” before
continuing.
23.
Run the complete system and test the pH of the final product. The pH does most likely need some
minor adjustment here to get the desired pH level.
24.
Use the scale on dosing unit (5) (closest to the system) for fine adjustments of pH level.
Do
not
change the setting on the first dosing unit at this point.
25.
Both dosing units’ adjustments can be locked after the final pH level has been reached.
Calibrating the level sensor
1.
Make sure the sensor is on. The green LED on the unit should be lid. The red LED indicates the signal
strength of the signal the unit sees. Under normal operation this let is on, if the red LED is not on the
sensor is not at the right angle or we are outside the 3-meter range.
2.
Install the sensor in the desired location. Make sure there are no edge the sensor can see. If the
sensor sees sharp edges the reading will be off.
3.
Push and hold the “analog button” for longer than 2 seconds.
Power LED = off, Output LED = on
4.
Hold a piece of cardboard at the position which equals 100% full tank.
5.
The red LED (signal) must be on of flashing
6.
‘Click” analog button
Power LED = off, Output LED = flashing
7.
Remove card board from 100% full and let the sensor see 0 % full
8.
“Click” analog button
9.
The red LED must be on or flashing
10.
If the operation was successful the Power and Output LED will be on
Pump start-up
Prime the pump before use. Open the inlet and outlet valves (if installed) to allow the pump to fill up with HOCl.
Open up some of the outlet to allow HOCl to “escape” the setup. Some pumps need to be primed by losing a vent
bolt. To find specifics on this type of priming please see pump manual. Turn on the pump and watch operation.
Make sure pump is running without excess noise.
System is now ready for operation.