4
Installation and start-up
Personnel:
n
Specialist
n
Qualified electrician
n
Mechanic
n
Service personnel
Protective equipment:
n
Protective gloves
n
Chemical-resistant protective gloves
n
Protective eyewear
n
Safety shoes
4.1
Installation work
4.1.1
Multiline 1000 and Multiline 1000 S
CAUTION!
The device has been pre-set to 230 V mains voltage! If the mains voltage is
115 V, rotate the voltage selector (pos. 1) by 180° and replace the fuse
(pos. 2) with the 115 V (250 mA) fuse included in the package.
1
2
3
4
Fig. 5: Setting the mains voltage
1
Voltage selector
2
Fuse
3
Cover plate with viewing window
Select voltage:
1.
Remove the cover plate (pos. 3).
2.
Turn the voltage selector insert (pos. 1) behind it by 180° so that the desired voltage
embossed on the insert is visible.
3.
Replace the cover plate (pos. 3). The set voltage is visible in the window of the cover
plate (pos. 3).
Installation and start-up
10
417102603 Rev. 1-02.2020
Содержание MULTILINE 1000