Eco Sensors
Ozone Analyzer, Model UV-100
11
Division of KWJ Engineering Inc.
User Manual, Rev. 2.0
[email protected] or (800) 472-6626
When the instrument is logging, the data output will be in the following form:
1,0.00,304.7,739.6,750,1.060,07/03/16,15:16:59
2,0.00,304.7,739.6,738,1.059,07/03/16,15:17:09
3,0.00,304.7,739.6,743,1.059,07/03/16,15:17:19
4,0.00,304.6,739.6,736,1.058,07/03/16,15:17:29
5,0.00,304.6,739.6,743,1.058,07/03/16,15:17:39
6,0.00,304.6,739.5,737,1.057,07/03/16,15:17:49
7,0.01,304.6,739.5,739,1.057,07/03/16,15:17:59
8,0.01,304.6,739.6,727,1.056,07/03/16,15:18:09
Logging Ended
where the columns of data are:
Sample Number, PPM, Cell Temperature (K), Cell Pressure (torr), Flow Rate, Photodiode Voltage, Date, Time
To download internally logged data, navigate to
Menu > Dat > Xmt
. This will print all the stored data
lines to the terminal program.
Logged Data
1,0.00,304.6,739.5,749,1.059,07/03/16,15:16:59
2,0.00,304.6,739.5,737,1.059,07/03/16,15:17:09
3,0.00,304.6,739.5,743,1.058,07/03/16,15:17:19
4,0.00,304.6,739.5,735,1.058,07/03/16,15:17:29
5,0.00,304.6,739.5,742,1.057,07/03/16,15:17:39
6,0.00,304.6,739.5,737,1.057,07/03/16,15:17:49
7,0.01,304.6,739.5,738,1.056,07/03/16,15:17:59
8,0.01,304.6,739.5,727,1.056,07/03/16,15:18:09
End of Log
where the columns of data are:
Sample Number, PPM, Cell Temperature (K), Cell Pressure (torr), Flow Rate, Photodiode Voltage, Date, Time
Different programs have different methods for saving or logging this data to a file. Please refer to the
user manual of the terminal program you are using for more details.
Accessing the Serial Menu
When the UV-100 is connected to a computer (via USB or serial cable), as described above,
commands can be sent to adjust instrument parameters and initiate logging tasks. Listed below are the
lower case command letters for performing operations while the instrument continues to measure.
l
Start logging and delete previously logged data
t
Transmit logged data
e
End logging
h
Output serial data line header
m
Serial menu